Activating the Onyx Vantage 9590 and Verifying Operation
The device contains numeric Light-Emitting Diodes (LEDs) that display oxygen saturation and pulse rate. A tricolor LED
display (pulse quality indicator, shown at left) provides a visual indication of the pulse signal quality, while blinking at the
corresponding pulse rate. This display changes colors to alert you to changes in pulse quality that may affect the readings:
Green indicates a good pulse signal.
Yellow indicates a marginal pulse signal.
Red indicates an inadequate pulse signal.
Activate the 9590 by inserting the patient’s finger into the device. The device detects the inserted finger and automatically
illuminates the displays. Correct positioning of the device on the finger is critical for accurate measurements.
NOTE: While on the finger, do not press the device against any surface and do not squeeze or hold it together. The internal
spring provides the correct pressure; additional pressure may cause inaccurate readings.
1. Insert the patient’s finger, nail side up, into the 9590 until the fingertip touches the built-in stop guide.
2. Make sure the finger is lying flat (not on its side) and is centered within the device. For best results, keep the device at
the patient’s heart or chest level.
3. If the device does not turn on, remove the finger and wait a few seconds before reinserting it.
When a finger is inserted, the device performs a brief startup sequence. Verify that all LEDs illuminate during the startup
sequence. If any LED is not lit, do not use the 9590; contact Nonin Technical Service for repair or replacement.
After the startup sequence, the device begins sensing the pulse (indicated by the blinking pulse quality indicator). Allow the
device to stabilize and observe about 4 seconds of continuous green-colored pulse quality before relying on the displayed
values. Continually verify operation. It is common for the displayed values to fluctuate slightly over a period of several
seconds. If the pulse quality indicator blinks yellow or red, try another finger.
A minus sign (-) appears in the left-most digit of the %SpO
2
display when the device senses the finger has been removed.
The last measured SpO
2
and pulse rate values display for 10 seconds while the device automatically turns off. The device
will automatically shut off (to conserve battery life) approximately 10 seconds after the finger is removed, or after a 2-minute
period of inadequate pulse signals.
If the 9590 does not turn on or if it shuts off unexpectedly:
Verify batteries are correctly inserted. Note: If batteries are installed backwards, the unit will not function.
The batteries are depleted. Replace batteries.
Continually verify operation. If the problem persists, remove the batteries and contact Nonin Technical Service.
The Oxitest
Plus7
by Datrend Systems, Inc. can be used to verify operation of the pulse oximeter.
Using the Lanyard
WARNING: Certain activities may pose a risk of injury, including strangulation, if lanyard should become
wrapped around your neck.
!
CAUTION: A flexible circuit connects the two halves. Do not twist or pull the flexible circuit or overextend the
device’s spring. Do not hang the lanyard from the device’s flexible circuit.
A lanyard is provided for convenience. The device will function with or without the lanyard.
If lanyard use is desired, thread the lanyard as shown below.
Pulse Quality
Indicator
%SpO
2
3


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nonin Onyx Vantage 9590 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nonin Onyx Vantage 9590 in the language / languages: English as an attachment in your email.

The manual is 2,48 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Nonin Onyx Vantage 9590

Nonin Onyx Vantage 9590 User Manual - German - 9 pages

Nonin Onyx Vantage 9590 User Manual - Dutch - 9 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info