Patient Alarms
Only one alarm condition can be reported to the VueLink at a time. Because of this,
patient alarm conditions are assigned priorities—with only the highest priority
alarm displayed at any one time. Alarms are sent to the VueLink even when the
oximeter’s alarms are silenced. Alarm priorities are assigned as follows:
1. SpO
2
Low Limit
2. Pulse Rate High Limit
3. Pulse Rate Low Limit
4. SpO
2
High Limit
5. Low Perfusion
Equipment Alarms
VueLink displays equipment alarms as “inoperatives.” Only one inoperative
condition can be reported to the VueLink at a time. Because of this, inoperatives are
assigned priorities—with only the highest priority displayed at any one time.
Inoperatives are sent to the VueLink even when the oximeter’s alarms are silenced.
Inoperative priorities are assigned as follows:
1. No SpO
2
Sensor
2. SpO
2
Sensor Fault
3. No SpO
2
Data
4. No Pulse Data
5. Low Battery
The following text strings, associated with the Avant 9600, are sent through the
VueLink module for display on the multiparameter monitor:
Parameter Displayed Text
Displayed High SpO2 Limit HiSpO2
Displayed Low SpO2 Limit LoSpO2
Displayed High Pulse Rate Limit Hi PR
Displayed Low Pulse Rate Limit Lo PR
SpO2 High Limit Alarm SpO2: HIGH
SpO2 Low Limit Alarm SpO2: LOW
Pulse Rate High Limit Alarm PULSE: HIGH
Pulse Rate Low Limit Alarm: PULSE: LOW
Low Perfusion Alarm LOW PERFUSION
Sensor Disconnect NO SpO2 SENSOR
Sensor Fault SpO2 SENSOR
No SpO2 Data Available NO SpO2 Data
No Pulse Data Available NO Pulse Data
Low Battery Battery: Pulse Ox
33


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nonin Avant 9600 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nonin Avant 9600 in the language / languages: English as an attachment in your email.

The manual is 0,7 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Nonin Avant 9600

Nonin Avant 9600 User Manual - German - 47 pages

Nonin Avant 9600 User Manual - Dutch - 25 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info