3.5. GENERAL SPECIFICATIONS
Working temperature range: - 10 °C … + 45 °C
Storage temperature range:
Max. humidity:
Pollution degree:
Protection classification:
- 20 … 70 °C
85 % RH (0 ÷ 40 °C)
2
double insulation
Over voltage category: Voltage inputs: CAT III 600 V
AC power supply CAT III 300 V
Protection degree: IP 44
Dimensions: 265 x 110 x 18.5 mm3
Weight (without accessories): 2 kg
3.6. MAINTENANCE
3.6.1. Batteries
Instrument contains rechargeable NiCd or NiMh batteries. Do NOT replace with
alkaline cells. These batteries should only be replaced with the same type as
defined on the battery cover label or in this manual.
Hazardous voltages exist inside this Instrument. Disconnect all test leads,
remove the power supply cable and switch off instrument before removing
battery compartment cover.
If it is necessary to replace batteries, all four MUST be replaced. Ensure batteries are
installed with the correct polarity; incorrect polarity can damage the batteries and/or
the Instrument.
There may exist special environmental regulations concerning the disposal of
batteries. These must be followed.
In case of blown battery fuse (F1), this should be replaced with the same type as
defined on the label close to it.
3.6.2. Cleaning
To clean the surface of instrument, use a soft cloth slightly moistened with soapy water or
alcohol. Then leave the instrument to dry totally before use.
Do not use liquids based on petrol or hydrocarbons!
Do not spill cleaning liquid over the instrument!
3.6.3. Periodic calibration
To ensure correct measurement, it is essential that the Instrument be regularly calibrated. If
used continuously on a daily basis, a six monthly calibration period is recommended,
otherwise annual calibration is sufficient.
16


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nieaf-Smitt PQA 6600 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nieaf-Smitt PQA 6600 in the language / languages: English as an attachment in your email.

The manual is 1.72 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info