8. Troubleshooting
Symptom Possible Cause Solution
Failed to
power on
Reading is
too low
Reading is
too high
Hi
Lo
Err
The battery level is too low
The infrared sensor lens of
the probe is dirty
The distance between the
probe and the target is too
great
You have just come from a
cold environment
You have just come from a hot
environment
You are exposed to hot air
generated by an air
conditioner or heater
Low battery
Replace the batteries.
Stay in a room between 10.0‒40.0
℃ (50.0‒104.0℉) for at least 30
minutes before taking a
temperature measurement.
Stay in a room between 10.0‒40.0
℃ (50.0‒104.0℉) for at least 30
minutes before taking a
temperature measurement.
Leave the environment for at
least 30 minutes before taking a
temperature measurement.
1. First make sure that the probe is clean
and free from dirt.
2. If the environmental temperature is out
of range, then both patient and
thermometer should stay in a room
between 10.0‒40.0℃ (50.0‒104.0℉) for at
least 30 minutes before taking a
temperature measurement.
3. If the problem persists after trying the
above method, please contact the dealer
or manufacturer.
1. First make sure that the probe is clean
and free from dirt.
2. If the environmental temperature is out
of range, then both patient and
thermometer should stay in a room
between 10.0‒40.0℃ (50.0‒104.0℉) for at
least 30 minutes before taking a
temperature measurement.
3. If the problem persists after trying the
above method, please contact the dealer
or manufacturer.
1. First make sure that the probe is clean
and free from dirt.
2. If the environmental temperature is out
of range, then both patient and
thermometer should stay in a room
between 10.0‒40.0℃ (50.0‒104.0℉) for at
least 30 minutes before taking a
temperature measurement.
3. If the problem persists after trying the
above method, please contact the dealer
or manufacturer.
Human body
temperature over
42.9℃/109.2℉
Human body
temperature below
32.0℃/89.6℉.
Environmental
temperature is not in
the 10.0‒40.0℃
(50.0‒104.0℉) range.
Let the probe cover touch
your forehead, or put the
probe into your ear canal.
Use cotton swabs dipped in
anhydrous ethanol to clean
the infrared sensor lens.
The polarities of the
batteries are reversed
The thermometer is damaged
Contact the dealer or manufacturer.
Ensure that the batteries are
inserted correctly.
Replace the batteries.
- 16 -
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for MostEssential PC809V at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of MostEssential PC809V in the language / languages: English as an attachment in your email.

The manual is 0,2 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info