783837
94
Zoom out
Zoom in
Previous page
1/100
Next page
D-2
D.2 Technical Alarm Messages
Alarm messages L Cause and action
SpO2 Self Test Err L The pulse oximeter failed the power-on self-test. Restart
the pulse oximeter.
SpO2 Comm Err L The pulse oximeter encountered a communication error.
Restart the pulse oximeter.
SpO2 Comm Stop L An error occurred to the SpO2 module or there was a
communication problem. Restart the pulse oximeter.
SpO2 Overrange L Measured SpO2 value is beyond the specified
measurement range. Check the patient’s condition.
PR Overrange L Measured PR value is beyond the specified measurement
range. Check the patient’s condition.
Sensor Off L*
No Sensor L
The SpO2 sensor detached the patient or the pulse
oximeter, or there was a fault with the SpO2 sensor, or an
unspecified SpO2 sensor was used. Check that the sensor
application site and the sensor type are correct, and make
sure that the sensor is undamaged. Reconnect the sensor if
the sensor is disconnected or use a new sensor if the
sensor is damaged.
SpO2 Low Perf L The signal detected is weak or the signal quality is poor.
Check the patient’s condition. Change the sensor
application site. If the problem persists, replace the sensor.
Voltage Too High L
Voltage Too Low L
The system power supply fails. Restart the pulse oximeter.
Battery Too Low M The battery power is low. Replace the batteries if alkaline
batteries are used or charge the battery if a lithium ion
battery is used.
Battery Error L A problem occurs when the lithium ion battery is being
charged. Check the battery for damage. If yes, replace the
battery.
Power Comm Err M Communication problem occurred to the power supply
part during the power-on self-test or operation. Restart the
pulse oximeter.
Clock Need Set L The real-time clock is reset. Set the clock.
94


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Mindray PM-60 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Mindray PM-60 in the language / languages: English as an attachment in your email.

The manual is 0.79 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Mindray PM-60

Mindray PM-60 User Manual - German - 104 pages

Mindray PM-60 User Manual - French - 102 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info