536876
14
Zoom out
Zoom in
Previous page
1/58
Next page
PF 100 12
9. Analysing Data using Computer Interface
The monitor can be connected via a USB connection port with a computer and all
memory data can be analysed by the Microlife Asthma Monitor software program.
Please contact your Microlife dealer.
10. Messages / Malfunctions / Errors
If an error occurs one of the following listed error codes is displayed.
Message Cause Remedy
Er2 Data communication between device
and printer does not work.
Check the cable connection
between device and printer.
no There is no data stored in the Monitor.
Hi The result is higher than 900 ml/min. This is a very good result.
Other possible malfunctions
If problems occur when using the device, the following points should be checked:
Malfunction Remedy
The display remains empty. 1. Check the polarity of the batteries (+/-).
Batteries have been
installed.
2. If the display is erratic orunusual, remove the
batteries and re-install new batteries.
The instrument frequently
fails to measure or the
values measured are
wrong.
1. Ensure that the measuring tube is connected
correctly.
2. Check if when blowing into the tube, the wing wheel
is rotating. Any objects, dust liquids or mucous may
interfere the rotation of the wing wheel. In this case
clean the tube as explained.
3. Discuss the values with your doctor.
If you have any questions regarding the use of this device, please ask your dealer
or pharmacist for the Microlife Service representative in your country. The Microlife
service team will be happy to help you.
11. Battery Replacement
When the battery low symbol appears in the display, the device is blocked until the
batteries have been replaced. Please use 1.5 V «Long-Life» or «Alkaline» batteries,
size «AAA». The use of 1.2 V Accumulator batteries is NOT recommended.
If the monitor is left unused for longer periods, please remove the batteries.
Please note that during battery exchange the data stored in the memory of the
monitor is well protected and will NOT be lost.
After battery change (or when the unit has been disconnected from any
power supply) time/date needs to be entered, again! Otherwise data can not
be memorised.
14


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife PF 100 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife PF 100 in the language / languages: English, German, French, Italian as an attachment in your email.

The manual is 2.5 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Microlife PF 100

Microlife PF 100 User Manual - English - 66 pages

Microlife PF 100 User Manual - German, French - 66 pages

Microlife PF 100 User Manual - Polish - 66 pages

Microlife PF 100 User Manual - Spanish - 66 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info