The device must be able to measure the pulse
properly to obtain an accurate SpO2 measure-
ment. Verify that nothing is hindering the pulse
measurement before relying on the SpO2
measurement.
The maximum application time at a single site
should be less than 30 minutes, in order to
ensure correct sensor alignment and skin integ-
rity.
Inaccurate measurements may occur if:
The patient suffers from significant levels of dysfunc-
tional hemoglobin (such as carboxyhemoglobin or
methemoglobin).
Intravascular dyes such as indocyanine green or meth-
ylene blue have been injected into the patient.
Used in the presence of high ambient light (e.g. direct
sunlight). Shield the sensor area with a surgical towel if
necessary.
There is excessive patient movement.
The patient experiences venous pulsations.
The patient has hypotension, severe vasoconstriction,
severe anemia, or hypothermia.
The patient is in cardiac arrest or is in shock.
Fingernail polish or false fingernails are applied.
6. Inserting the batteries
6
After you have unpacked your device, first insert the
batteries. The battery compartment is on the bottom of the
device. Remove the battery cover by sliding it in the direc-
tion shown. Insert the batteries (2 x 1.5 V, size AAA),
thereby observing the indicated polarity.
Replace the batteries when the low power indi-
cator 5 appears on the display.
Always replace both batteries at the same time.
7. ON/OFF button
1
/ Function button
Press and release the ON/OFF button 1 to turn on, hold
the button for about one second. The device shows the
settings menu AT. Press or hold the ON/OFF button to
perform corresponding operations. Hold it to set an item, or
press it to switch an option or switch the display mode.
Press means no more than 0.5 seconds, while hold means
more than 0.5 seconds.
Alert Sound Setting
Hold ON/OFF button 1 while the device is switched
on.Settings menu (interface 1) is displayed AT-a. Move «*»
to the corresponding option, and hold the function button to
set Alm to on and set Beep to off. When Alm is set to on
and the measured values of the blood oxygen saturation
and pulse rate go beyond the upper limit or lower limit, the
device gives off an alert sound. When Beep is set to on, a
tick will be heard along with pulse beats during pulse rate
measurement. While the «*» symbol stays on the Restore
option, hold the functional button to restore default settings.
Brightness Setting
Press ON/OFF button 1 to select the Brightness option
and then hold the ON/OFF button to set the brightness to a
value ranging from 1 to 5. The greater the value, the
greater the brightness of the screen.
Alert Range Setting
On settings menu (interface 2) AT-b, press the ON/OFF
button 1 to switch between options. On this interface, you
can set the upper limit and lower limit of SpO2 Alm and PR
Alm. While the «*» symbol stays on the +/- option, hold the
functional button to set the option to + or -.
In + mode, select the corresponding option and hold the
ON/OFF button to increase the upper or lower limit; in -
mode, hold the ON/OFF button to decrease the upper or
lower limit. Move «*» to the Exit option, and hold the ON/
OFF button to return to the monitoring interface.
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife OXY 210 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife OXY 210 in the language / languages: English, German, Dutch, French, Italian, Portuguese, Turkish, Spanish as an attachment in your email.

The manual is 1.07 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info