773464
6
Zoom out
Zoom in
Previous page
1/81
Next page
5
In the following situations it is recommended that three temper-
atures in the same ear be taken and the highest one taken as
the reading:
1. New born infants in the first 100 days.
2. Children under three years of age with a compromised
immune system and for whom the presence or absence of
fever is critical.
3. When the user is learning how to use the thermometer for
the first time until he/she has familiarized himself/herself with
the device and obtains consistent readings.
4. If the measurement is surprisingly low.
Don't take a measurement while or immediately after nursing a
baby.
Don't use the thermometer in high humidity environments.
Patients should not drink, eat or exercise before/while taking
the measurement.
Doctors recommend rectal measurement for newborn infants
within the first 6 months, as all other measuring methods might
lead to ambiguous results.
Readings from different measuring sites should not be
compared as the normal body temperature varies by
measuring site and time of day, being highest in the evening
and lowest about one hour before waking up.
Normal body temperature ranges:
- Axillar: 34.7 - 37.3 °C / 94.5 - 99.1 °F
- Oral: 35.5 - 37.5 °C / 95.9 - 99.5 °F
- Rectal: 36.6 - 38.0 °C / 97.9 - 100.4 °F
- Microlife IR 210: 35.4 - 37.4 °C / 95.7 - 99.3 °F
8. Changing between Celsius and Fahrenheit
This thermometer can display temperature readings in either Fahr-
enheit or Celsius. To switch the display between °C and °F, press
and hold the MODE button 4 for 3 seconds; the beeper icon is
shown on the display. Press the MODE button again; the current
measurement scale («°C» or «°F» icon) will be shown on the
display AQ. Change the measurement scale between °C and °F by
pressing the M-button 3. When the measurement scale has been
chosen, press the START/IO button 6 to enter the «ready for
measuring» mode; otherwise the device automatically switches to
ready for measuring after 10 seconds 9.
9. How to recall 30 readings in Memory Mode
This thermometer can recall the last 30 readings with a record of
both time and date.
Recall mode AR: Press the M-button 3 to enter recall mode
when the power is off. The memory icon «M» will flash.
Reading 1 - the last reading AS: Press and release the M-
button 3 to recall the last reading. Number «1» and a flashing
«M» are displayed.
Pressing and releasing the M-button
3
after the last 30 readings
have been recalled will resume the above sequence from reading 1.
10.Error Messages
Measured temperature too high BT:Displays «H» when
measured temperature is higher than 43 °C / 109.4 °F in body
mode or 99.9 °C / 211.8 °F in object mode.
Measured temperature too low BK:Displays «L» when
measured temperature is lower than 32 °C / 89.6 °F in body
mode or 0,1 °C / 32,2 °F in object mode.
Ambient temperature too high BL: Displays «AH» when
ambient temperature is higher than 40.0 °C / 104.0 °F.
Ambient temperature too low BM: Display «AL» when
ambient temperature is lower than 10.0 °C / 50.0 °F.
Incorrect location indication AN: The probe is not correctly
inserted in the ear canal. Please insert the probe as described
in this manual.
Error function display AO: The system has a malfunction.
Blank display BO: Check if the battery has been loaded
correctly. Also check polarity (<+> and <->) of the battery.
Flat battery indication BP: If only the «battery» icon is shown
on the display, the battery should be replaced immediately.
11.Cleaning and Disinfecting
Use an alcohol swab or cotton tissue moistened with alcohol (70%
Isopropyl) to clean the thermometer casing and the measuring
sensor. Ensure that no liquid enters the interior of the device. Never
use abrasive cleaning agents, thinners or benzene for cleaning and
never immerse the device in water or other cleaning liquids. Take
care not to scratch the surface of the sensor lens and the display.
12.Battery Replacement
This device is supplied with one lithium battery, type CR2032. The
battery needs replacing when only the «battery» icon BP is shown
on the display.
Remove the battery cover BS by sliding it in the direction shown.
Insert the new battery with the + at the top.
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife IR 210 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife IR 210 in the language / languages: English, German, Dutch, French, Italian, Portuguese, Turkish, Spanish as an attachment in your email.

The manual is 7.46 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info