Guidance and manufacturer's declaration - electromagnetic immunity
The BP4GAPO-2M is intended for use in the electromagnetic environment specied below. The customer or the user of the BP4GAPO-2M should assure that
it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment-guidance
Electrostatic discharge
(ESD)
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood, concrete or ceramic tile. If oors are covered
with synthetic material, the relative humidity should be at least 30 %
Electrical fast transient/burst
IEC 61000-4-4
+ 2kV for power supply lines
+ 1kV for input/output lines
Not applicable
Not applicable
Mains power quality should be that of a typical commerical or
hospital environment
Surge IEC 61000-4-5 + 1kV line(s) to line(s)
+ 2kV line(s) to earth
Not applicable
Not applicable
Mains power quality should be that of a typical commerical or
hospital environment
Voltage Dips, short
interruption and voltage
variations on power supply
input lines IEC 61000-4-11
<5% UT (>95% dip in UT) for 0,5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 s
Not applicable
Not applicable
Not applicable
Not applicable
Mains power quality should be that of a typical commerical or
hospital environment. If the user of the BP4GAPO-2M requires
continued operation during power mains interruptions, it is
recommended that the BP4GAPO-2M be powered from an
uninterruptible power supply or a battery.
Power frequency (50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m The BP4GAPO-2M power frequency magnetic elds should be at
levels characteristic of a typical location in a typical commercial or
hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
18


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife Cardio plus at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife Cardio plus in the language / languages: English as an attachment in your email.

The manual is 0,68 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info