704042
6
Zoom out
Zoom in
Previous page
1/94
Next page
4
patients first visit in order to determine which arm to measure in the
future. The arm with the higher blood pressure should be measured.
Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do not
interfere with the cuff if they are laid flat.
Always ensure that the correct cuff size is used (marking on the cuff).
Fit the cuff closely, but not too tight.
Make sure that the cuff is positioned 2 cm above the elbow.
The artery mark on the cuff (ca. 3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
Support your arm so it is relaxed.
Ensure that the cuff is at the same height as your heart.
4. Taking a Blood Pressure Measurement
1. Select standard (single measurement) or MAM mode (automatic
triple measurement): see details in chapter «2.».
2. Press the ON/OFF button 1 to start the measurement.
3. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
4. The cuff fit check AR on the display indicates that the cuff is
perfectly placed. If the icon AR-A appears, the cuff is fitted
suboptimally, but it is still ok to measure.
5. When the correct pressure is reached, the pumping stops and the
pressure falls gradually. If the required pressure was not reached,
the device will automatically pump some more air into the cuff.
6. During the measurement, the pulse indicator BM flashes in the
display.
7. The result, comprising the systolic AL and the diastolic AM blood
pressure and the pulse rate AN is displayed. Note also the
explanations on further display symbols in this booklet.
8. When the device has finished measuring, remove the cuff.
9. Switch off the device. (The monitor does switch off automatically
after approx. 1 min.).
AF detection is only activated in MAM mode.
You can stop the measurement at any time by pressing the
ON/OFF button (e.g. if you feel uneasy or an unpleasant
pressure sensation).
This monitor is specially tested for use in pregnancy and
pre-eclampsia. When you detect unusual high readings in
pregnancy, you should measure after a short while again
(eg. 1 hour). If the reading is still too high, consult your
doctor or gynecologist.
In pregnancy the AFIB symbol can be ignored.
Manual inflation
In case of high systolic blood pressure (e.g. above 135 mmHg)
,
it can be an advantage to set the pressure individually. Press the ON/
OFF button after the monitor has been pumped up to a level of
approx. 30 mmHg (shown on the display). Keep the button pressed
until the pressure is about 40 mmHg above the expected systolic
value – then release the button.
How not to store a reading
As soon as the reading is displayed press and hold the ON/OFF
button 1 until «M» AP is flashing. Confirm to delete the reading by
pressing the MAM button 4.
«CL» is displayed when the reading is deleted from the
memory successfully.
How do I evaluate my blood pressure?
The triangle on the left-hand edge of the display
BL
points at the
range within which the measured blood pressure value lies. The
value is either within the optimum (green), elevated (yellow) or high
(red) range. The classification corresponds to the following ranges
defined by international guidelines (
ESH, ESC, JSH
). Data in mmHg.
The higher value is the one that determines the evaluation.
Example: a blood pressure value of 140/80 mmHg or a value of
130/90 mmHg indicates «blood pressure too high».
Appearance of the Pulse Arrhythmia (PAD)
This symbol BT indicates that certain pulse irregularities were
detected during the measurement. In this case, the result may
deviate from your normal blood pressure – repeat the measure-
ment. In most cases, this is no cause for concern. However, if the
symbol appears on a regular basis (e.g. several times a week with
measurements taken daily) we advise you to tell your doctor.
Please show your doctor the following explanation:
Range Systolic Diastolic Recommendation
1. blood pressure
too high
135 85 Seek medical
advice
2. blood pressure
elevated
130 - 134 80 - 84 Self-check
3. blood pressure
normal
<130 <80 Self-check
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife BP B3 AFIB - 2019 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife BP B3 AFIB - 2019 in the language / languages: English, Polish as an attachment in your email.

The manual is 8,66 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Microlife BP B3 AFIB - 2019

Microlife BP B3 AFIB - 2019 User Manual - English, German, Dutch, French, Italian, Portuguese, Turkish, Spanish - 93 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info