5. Always ensure that the correct cuff size is used (marking on the
cuff).
Fit the cuff closely, but not too tight.
Make sure that the cuff is positioned 1-2 cm above the
elbow.
The artery mark on the cuff (ca. 3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
Support your arm so it is relaxed.
Ensure that the cuff is at the same height as your heart.
6. Any improper setup or damage to the stethoscope will cause
distorted sound or poor sound transmission causing inaccurate
readings.
7.
Proper deflation rate is essential for an accurate reading.
Practice and master a recommended deflation rate of
2-3
mmHg
per second or a drop of
1-2
marks on the manometer
1
for each
heartbeat.
Measuring procedure
1. Place the chest piece underneath the cuff 2 or 1-2 cm below
it. Make sure the chest piece is in contact with the skin and lies
on the brachial artery.
2. Plug in the ear piece and check, if the chest piece is placed
correctly, so that the Korotkoff sounds appear the loudest.
3. Close the valve 4 on the pump ball 3 by turning the screw
clockwise. Do not over-tighten.
4. Take the pump ball
3
in your free hand (the arm you are not
using to measure) and pump up the cuff. Watch the pressure indi-
cator on the manometer
1
and pump up to approx. 40 mmHg
higher than the expected systolic value (the upper value).
Inflate to 200 mmHg if you are not sure about the expected
value.
5. Open the valve 4 slowly by turning the screw counter clock-
wise whilst holding the stethoscope chest piece on the brachial
artery. Listen carefully as the cuff begins to deflate. Note the
reading on the manometer 1 as soon as you hear a faint,
rhythmic tapping or thumping sounds. This is the systolic
blood pressure reading.
6. Allow the pressure to continue dropping at the same deflation
rate. Note the reading on the manometer 1 as soon as the
thumping sound stops. This is the diastolic blood pressure
reading.
7. Deflate the cuff completely.
8. Repeat the measurement at least two further times and record
your values, date and time immediately after finishing the
measurements.
9. Remove the cuff and the stethoscope.
4. Malfunction / Troubleshooting
If problems occur when using the device, the following points
should be checked and if necessary, the corresponding measures
are to be taken:
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
Description Potential cause and remedy
The sound transmis-
sion is poor, distorted
or there is extraneous
noise.
Check the ear pieces if they are dirty
or cracked. If not, make sure you
wear them properly.
Check the tube if it is broken or
twisted.
Check the chest piece if there is any
damage.
Make sure the chest piece is in
contact with the skin and lies on the
brachial artery. Clean or replace any
defective parts if found to avoid inac-
curate reading.
The pressure does not
rise although the pump
ball 3 is pumping.
Make sure that the valve is closed.
Make sure the cuff is properly
connected to bulb and manometer.
Check if the cuff, tube and/or bulb is
leaking. Replace the defective parts
if any.
The deflation rate can
not be set to 2-
3 mmHg/sec. by
adjusting the valve 4.
Disassemble the valve from pump ball
to check if there is any blockage in the
airway of the valve. Clean the blockage
and try again. If it still does not work,
replace it to avoid inaccurate readings.
The manometer needle
is not at 0 ± 3 mmHg at
rest.
Make sure that the valve is
completely open for zero check.
If still more than 3 mmHg deviation,
contact your dealer to recalibrate the
manometer.
5


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife BP AG1-10 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife BP AG1-10 in the language / languages: English, French, Italian as an attachment in your email.

The manual is 0.72 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info