EN
How do I evaluate my blood pressure?
The higher value is the one that determines the evaluation.
Example: a readout value between 150/85 or 120/98 mmHg indi-
cates «blood pressure too high».
The slot-in card 3 on the front of the instrument shows ranges
1-6 in the Table.
2. Using the Instrument for the First Time
Activating the fitted batteries
Pull out the protective strip projecting from the battery compart-
ment 7.
Selecting the correct cuff
Microlife offers 3 different cuff sizes: S, M and L. Select the cuff
size to match the circumference of your upper arms (measured by
close fitting in the centre of the upper arm). M is the correct size
for most people.
Connecting the pump ball
Connect the pump ball AT to the instrument by firmly inserting the
connector AK into the pump ball socket 5.
3. Taking a Blood Pressure Measurement using this
Instrument
Table for classifying blood pressure values in adults in accordance
with the World Health Organisation (WHO) in 2003. Data in mmHg.
Range Systolic Diastolic Recommendation
blood pressure too low z 100 z 60 Consult your doctor
1. blood pressure optimum 100 - 120 60 - 80 Self-check
2. blood pressure normal 120 - 130 80 - 85 Self-check
3. blood pressure slightly
high
130 - 140 85 - 90 Consult your doctor
4. blood pressure too high 140 - 160 90 - 100 Seek medical advice
5. blood pressure far too
high
160 - 180 100 - 110 Seek medical advice
6. blood pressure danger-
ously high
180 y 110 y Urgently seek
medical advice!
Cuff size for circumference of upper arm
S 17 - 22 cm (6.75 - 8.75 inches)
M 22 - 32 cm (8.75 - 12.5 inches)
L 32 - 42 cm (12.5 - 16.5 inches)
)
Only use Microlife cuffs.
X
Contact Microlife Service if the enclosed cuff 8 does not fit.
X
Connect the cuff to the instrument by inserting the cuff
connector 9 into the cuff socket 4 as far as it will go.
Checklist for taking a reliable measurement
1. Avoid activity, eating or smoking immediately before the
measurement.
2. Sit down for at least 5 minutes before the measurement and
relax.
3. Always measure on the same arm (normally left).
4. Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do
not interfere with the cuff if they are laid flat.
5. Always ensure that the cuff is positioned correctly, as shown
in the pictures illustrated on the short instruction card.
Fit the cuff closely, but not too tight.
Make sure that the cuff is 2 cm (0.75 inch) above your
elbow with the tube on the inside of your arm.
Support your arm so it is relaxed.
Ensure that the cuff is at the same height as your heart.
6. Switch on the instrument and wait until «0» appears in the
display and the arrow «S» BT starts to flash.
7. Take the pump ball in your free hand (the arm you are not
measuring from) and pump up the cuff. Watch the pressure
indication in the display and pump approx. 40 mmHg higher
than the expected systolic value (the upper value). If you
have not pumped enough, a flashing arrow «S» BT will
appear telling you to pump higher.
8.
After pumping, the measurement is taken automatically. Relax,
do not move and do not tense your arm muscles until the mea-
surement result is displayed. Breathe normally and do not talk.
9.
During the measurement, the heart symbol
AN
flashes in the
display and a beep sounds every time a heartbeat is detected.
10. The result, comprising the systolic AQ and the diastolic AR
blood pressure and the pulse AS is displayed and longer beep
is heard. Note also the explanations on further displays in
this booklet.
5


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife BP A80 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife BP A80 in the language / languages: English, German, Dutch, French, Polish, Portuguese, Turkish, Spanish as an attachment in your email.

The manual is 4,52 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info