9. Technical specifications
Weight: 575 g (with batteries + cuff size M)
Size: 106 (W) x 162 (L) x 61.5 (H) mm
Storage temperature: –5 to +50 °C
Humidity: 15 to 90% relative humidity maximum
Operation temperature: 10 to 40 °C
Display: LCD-Display (Liquid Crystal Display)
Measuring method: oscillometric
Pressure sensor: capacitive
Measuring range:
SYS/DIA: 30 to 280 mmHg
Pulse: 40 to 200 per minute
Cuff pressure display range: 0–299 mmHg
Memory: Storing the last measurements automatically
Measuring resolution: 1 mmHg
Accuracy:
Pressure: within ± 3 mmHg
Pulse: ± 5 % of the reading
Power source: a) 4 x 1.5 V Batteries; size AA
b) Mains adapter DC 6V, 600 mA (optional)
Cuffs: type M for arm circumference 22–32 cm or
type L for arm circumference 32–42 cm (optional)
Reference to standards: EU Directives 93/42/EEC; NIBP requirements:
EN 1060-1 /-3 /-4, ANSI / AAMI SP10
Technical alterations reserved!
10. www.microlife.com
Detailed user information about our thermometers and blood pressure monitors as well as services can be
found at www.microlife.com.
11