5. Safety, Care, Accuracy Test and Disposal
Safety and protection
This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Never inflate beyond 300 mmHg.
Always deflate the cuff completely before storage.
Do not use this device if you think it is damaged or notice
anything unusual.
Never open this device.
Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed.
Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soapsuds.
WARNING: Do not wash the cuff in a washing machine or
dishwasher!
Accuracy test
We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).
Disposal
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
6. Guarantee
This device is covered by a 2 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
Cuffs and parts that become worn with use are not included.
Opening or altering the device invalidates the guarantee.
The guarantee does not cover damage caused by improper
handling, accidents or non-compliance with the operating
instructions.
Please contact your local Microlife-Service (see foreword).
7. Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Weight:
450 g
Dimensions:
175 x 70 x 103
mm
Storage temperature:
-20 - +70 °C
85
% relative maximum humidity
Operating temperature:
0 - 46 °C
Measurement range:
0 - 300 mmHg
Resolution:
2 mmHg
Static accuracy:
within ± 3 mmHg between 18 - 33 °C;
within ± 6 mmHg between 34 - 46 °C
Air leakage: < ± 4 mmHg/min
Hysteresis error: within 0 - 4 mmHg
Included accessories: Cuff (22 - 32 cm), pump ball, valve,
stethoscope (chest piece attached to
the cuff), softbag
Reference to
standards:
EN 1060-1/-2; ANSI / AAMI SP09
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife AG1-30 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife AG1-30 in the language / languages: English, French, Italian as an attachment in your email.

The manual is 1,76 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info