5000-count portable digital multimeter, with current inputs
25
6. TECHNICAL SPECIFICATIONS
6.1. Genera
l
Only the values assigned tolerances or given limits are guaranteed values. Values without
tolerances are giv
en for information only, without guarantee (s
tandard NFC 42670). The
measurement errors must be considered in an env
ironment of reference temperature (refer to
§
6.2.11).
It is essential that all measuring instruments are regularly calibrated.
6.2. Characteristics
Accuracy is ± [% reading (R) + number of digits (D)].
{Accuracy : ‘’n %R + n D’ means ‘’n % of the reading + n digits’’ }
6.2.1. DC voltages
Selector switch
position
Ranges
Accuracy
Input impedance
Protection
(∗)
Resolution
ADPDC 500 mVDC
10 MΩ
0.1 mVDC
5 VDC
11 MΩ ± 1100 V
PEAK
1 mV
DC
VDC 50 VDC 0,3 % R + 2 D 775 Vrms 10 mVDC
500 VDC
10 MΩ
100 mVDC
1000 VDC 1 VDC
(∗) Acceptance permanent max. voltage
Number of points : 5 000
Range selection : automatic or manual for the 5 V, 50 V,
500 V, 1000 V ranges
Common mode rejection : at 50 and 60 Hz, better than 120 dB
Serial mode rejection : at 50 and 60 Hz, better than 60 dB
Intermittent signal sounds and « OL » on the display in case of range overpassing.
For measurements performed on alternative signals, the selected range must tally with the max.
value of the signal peak.
Additional error in Pk +/- mode for a pulse width of 1 ms : 1 % R ± 50 D
6.2.2.
AC voltages (AC and AC+DC)
Switch Range Input Protection Accuracy Resolution
position impedance (*) 40Hz to 1 kHz 1 to 4 kHz 4 to 20 kHz 20 to 50 kHz 50 to 100 kHz
5 VAC 1 mVAC
VLOW Z 50 VAC
500 kΩ
± 1100 Vpk 1 % L + 3 D 1 % L + 3 D 2 % L + 3 D 3 % L + 3 D 5 % L + 3 D 10 mVAC
500 VAC 775 Vrms 100 mVAC
750 VAC - - - - 1 mVAC
ADPAC+DC 500 mVAC+DC
10 MΩ
//100 pF
± 1100 Vpk
775 Vrms
1.5 % L + 3 D - - - - 0.1 mV
AC+DC
5 VAC+DC
11 MΩ
1 mVAC+DC
VAC+DC 50 VAC+DC ± 1100 Vpk 1 % L + 3 D 1 % L + 3 D 2 % L + 3 D 3 % L + 3 D 5 % L + 3 D 10 mVAC+DC
500 VAC+DC
10 MΩ
775 Vrms 100 mVAC+DC
750 VAC+DC - - - - 1 VAC+DC
(∗) Acceptance permanent max. voltage
Specifications applicable from 10 % to 100 % of the range for a sinusoidal signal in band 40 Hz
to 20 kHz and from 20 % to 100 % of the range in band 20 kHz to 100 kHz.
Number of points : 5 000
Range selection : automatic or manual for the 5 V, 50 V,
500 V, 750 V ranges
Common mode rejection : at 50 and 60 Hz, better than 60 dB
Additional error according to crest factor : 0.5 % for a crest factor of 2 to 3
1 % for a crest factor of 3 to 6
Intermittent signal sounds and « OL » on the display in case of range overpassing.
14


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrix MX 26 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrix MX 26 in the language / languages: English as an attachment in your email.

The manual is 0,22 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrix MX 26

Metrix MX 26 User Manual - German - 18 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info