MI 3100 s, MI 3100 SE EurotestEASI Instrument description
17
3
1
2
R: 0.0 1999 Test current: max. 8.5mA
Open-circuit voltage: 6.5V

Continuity 7mA



R: 0.18M 199.9M , U =50V ,100 ,
R: 0.12M 999M , U = 500V , 1kV
U: 0V 1200V
Nominal voltages: 100V , 250V , 500V , 1kV
Measuring current: min. 1mA at R =U 1k /V
Short-circuit current: < 3mA



N
N
NN


Insulation resistance (EN 61557-2)
V250V

Continuity
Tripping time
non-delayed (time-delayed) RCDs



1: 0ms 300ms (500ms)
2 : 0ms 150ms (200ms)
5 : 0ms 40ms (150ms), U : 0.0V 100.0V
C
Tripping current
I : 0.2 I 1.1 I AC (
t : 0ms 300ms, U : 0.0V 100.0V



C

NN N

1.5 I A)
Multiplier: 0.5, 1, 2, 5

Contact voltage
U : 0.0V 100.0V
R : 0. 00 10 .00k (R =U / I )
C
SSCN

R : 0. 1999 I : 0. A . kA
L-N(L) PSC
17 20 1 4
Nominal voltage: 100V 440V / 15Hz 500Hz


Line impedance (EN 61557-3)

Fault loop (EN 61557-3)
R : 0. 1999
I : 0. A .kA

PFC
14 1 4
Nominal voltage: 100V 264V / 15Hz 500Hz

impedance
L-PE
17

Voltage, frequency
U: 0V 440V / f: 15Hz 500Hz

Phase rotation (EN 61557-7)
Nominal voltage: 100V 440V / 1 00
Results: 1.2.3 or 2.1.3

5Hz 5 Hz
RCD (EN 61557-6)
I : 10mA, 30mA, 100mA, 300mA, 500mA, 1A
Nominal voltage: 100V 264V/ 15Hz 500Hz
 
R Low (EN 61557-4)


R: 0.12 1999
Test current: min. ±200mA at 2
Open-circuit voltage: 6.5V 9.0V

Resistance to earth (EN 61557-5)
R : 0.04 9999
Open-circuit voltage :
Short-circuit current

< 45V
: < 20mA
RMS
Ljubljanska 77
SI - 1354 Horjul
Tel: +386 1 75 58 200
http://www.metrel.si
20 224 832
CAT III 600V
550V
Figure 3.5: Bottom
Legend:
1 Bottom information label
2 Neck belt openings
3 Handling side covers
3.4 Carrying the instrument
With the neck-carrying belt supplied in standard set, various possibilities of carrying the
instrument are available. Operator can choose appropriate one on basis of his operation, see
the following examples:
The instrument hangs
around operators neck
only – quick placing and
displacing.
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI3100S at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI3100S in the language / languages: English as an attachment in your email.

The manual is 1.81 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI3100S

Metrel MI3100S User Manual - German - 92 pages

Metrel MI3100S User Manual - Dutch - 25 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info