MI 3100 s, MI 3100 SE EurotestEASI Safety and operational considerations
8
Warnings related to safety of measurement functions:
Insulation resistance
Insulation resistance measurement should only be performed on de-energized objects!
Do not touch the test object during the measurement or before it is fully discharged! Risk
of electric shock!
When an insulation resistance measurement has been performed on a capacitive object,
automatic discharge may not be done immediately! The warning message and the
actual voltage are displayed during discharge until voltage drops below 30 V.
Do not connect test terminals to external voltage higher than 600 V (AC or DC) in order
not to damage the test instrument!
Continuity functions
Continuity measurements should only be performed on de-energized objects!
Parallel loops may influence on test results.
Testing PE terminal
If phase voltage is detected on the tested PE terminal, stop all measurements
immediately and ensure the cause of the fault is eliminated before proceeding with any
activity!
Notes related to measurement functions:
General
The indicator means that the selected measurement cannot be performed because
of irregular conditions on input terminals.
Insulation resistance, continuity functions and earth resistance measurements can only
be performed on de-energized objects.
PASS / FAIL indication is enabled when limit is set. Apply appropriate limit value for
evaluation of measurement results.
In the case that only two of the three wires are connected to the electrical installation
under test, only voltage indication between these two wires is valid.
Insulation resistance
The standard three-wire test lead, schuko test cable or Plug / Tip commanders can be
used for the insulation test with voltages 1kV.
If a voltage of higher than 30 V (AC or DC) is detected between test terminals, the
insulation resistance measurement will not be performed.
The instrument automatically discharge tested object after finished measurement.
A double click of TEST key starts a continuous measurement.
Continuity functions
If a voltage of higher than 10 V (AC or DC) is detected between test terminals, the
continuity resistance test will not be performed.
Compensate test lead resistance before performing a continuity measurement, where
necessary.
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI3100S at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI3100S in the language / languages: English as an attachment in your email.

The manual is 1.81 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI3100S

Metrel MI3100S User Manual - German - 92 pages

Metrel MI3100S User Manual - Dutch - 25 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info