MI 3152(H) EurotestXC (2,5 kV) Instrument operation
28
Instrument is overheated. The measurement is prohibited until the
temperature decreases under the allowed limit.
High electrical noise was detected during measurement. Results may be
impaired.
Indication of noise voltage above 5 V between H and E terminals during earth
resistance measurement.
L and N are changed.
In most instrument profiles L and N test terminals are reversed automatically
according to detected voltages on input terminal. In instrument profiles for
countries where the position of phase and neutral connector is defined the
selected feature is not working.
Warning! High voltage is applied to the test terminals.
The instrument automatically discharge tested object after finished insulation
measurement.
When an insulation resistance measurement has been performed on a
capacitive object, automatic discharge may not be done immediately! The
warning symbol and the actual voltage are displayed during discharge until
voltage drops below 30 V.
Warning! Dangerous voltage on the PE terminal! Stop the activity
immediately and eliminate the fault / connection problem before proceeding
with any activity!
Continuous sound warning and yellow coloured screen is also present.
Test leads resistance in R low / Continuity measurement is not compensated.
Test leads resistance in R low / Continuity measurement is compensated.
High resistance to earth of current test probes. Results may be impaired.
High resistance to earth of potential test probes. Results may be impaired.
High resistance to earth of potential and current test probes. Results may be
impaired.
Too small current for declared accuracy. Results may be impaired. Check in
Current Clamp Settings if sensitivity of current clamp can be increased.
In Earth 2 Clamp measurement results are very accurate for resistances
below 10 Ω. At higher values (several 10 Ω) the test current drops to few mA.
The measuring accuracy for small currents and immunity against noise
currents must be considered!
Measured signal is out of range (clipped). Results are impaired.
Single fault condition in IT system. (MI 3152 only)
Fuse F1 is broken.
29


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3152 - EurotestXC at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3152 - EurotestXC in the language / languages: English as an attachment in your email.

The manual is 6.49 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3152 - EurotestXC

Metrel MI 3152 - EurotestXC User Manual - German - 213 pages

Metrel MI 3152 - EurotestXC User Manual - Dutch - 85 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info