MI 3109 EurotestPV Lite Technische Spezifikationen
61
9.2 Durchgang
9.2.1 Widerstand R LOW
Der Messbereich gemäß EN61557 ist 0,16 1999 .
Messbereich R ()
Auflösung ()
Genauigkeit
0,00 19,99
0,01
(3 % des Ablesewerts + 3 Ziffern)
20,0 199,9
0,1
(5 % des Ablesewerts)
200 1999
1
Leerlaufspannung ............................. 6,5 V DC 9 V DC
Messstrom ......................................... min. 200 mA bei Lastwiderstand von 2
Kompensation der Prüfleitung ........... bis zu 5
Anzahl möglicher Prüfungen .............. > 2000, bei vollständig geladener Batterie/Akku
Automatische Polaritätsumkehrung der Prüfspannung.
9.2.2 Widerstand CONTINUITY
Messbereich ()
Auflösung ()
Genauigkeit
0,0 19,9
0,1
(5 % des Ablesewerts + 3 Ziffern)
20 1999
1
Leerlaufspannung ............................. 6,5 V DC 9 V DC
Kurzschlussstrom .............................. max. 8.5 mA
Kompensation der Prüfleitung ........... bis zu 5
9.3 PV-Prüfungen
9.3.1 Genauigkeit der STC-Daten
Die Genauigkeit der STC-Werte basiert auf der Genauigkeit der gemessenen
elektrischen Größen, der Genauigkeit der Umgebungsparameter und der eingegebenen
Parameters für das PV-Modul. Siehe Anhang E: PV-Messungen Berechnungswerte,
sollten Sie weitere Informationen zur Berechnung der STC-Werte benötigen.
9.3.2 Modul, Wechselrichter
DC-Spannung
Messbereich (V)
Auflösung (V)
Genauigkeit
0,0 14.9
0,1
indicativ
15,0 199,9
0,1
(1,5 % des Ablesewerts + 5 Ziffern)
200 999
1
1,5 % des Ablesewerts
DC-Strom
Messbereich (A)
Auflösung (mA)
Genauigkeit
0,00 19,99
10
(1,5 % des Ablesewerts + 5 Ziffern)
20,0 199,9
100
1,5 % des Ablesewerts
200 299
1000
1,5 % des Ablesewerts
61


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3109 - Eurotest PV Lite at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3109 - Eurotest PV Lite in the language / languages: Dutch as an attachment in your email.

The manual is 2.13 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3109 - Eurotest PV Lite

Metrel MI 3109 - Eurotest PV Lite User Manual - English - 74 pages

Metrel MI 3109 - Eurotest PV Lite User Manual - Dutch - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info