MI 3109 EurotestPV Lite Instrument operation
26
Keys:
/ Selects option.
TEST Enters menu for changing parameters.
ESC Exits back to settings menu.
Function selector Exits back to main menu without changes.
PV module settings
Parameters of PV modules can be set in this menu. A database for up to 20 PV
modules can be created / edited. Parameters are used for calculation of STC values.
Note:
The database can be also created on the PC or mobile device and then sent to
the instrument. PCSW EurolinkPRO and some Android applications support this
feature.
Figure 4.13: PV module settings menu
Parameters of PV module:
Module PV module name
Pmax 1 W ..2000 W Nominal power of PV module
Umpp 1.0 V .. 999 V Voltage on maximum power point
Impp 0.01 A .. 15.0 A Current on maximum power point
Uoc 1.0 V .. 999 V Open circuit voltage of module
Isc 0.01 A .. 15.0 A Short circuit current of module
NOCT 1.0 °C .. 99.0 °C Nominal working temperature of PV cell
alfa -5.00 mA/°C .. 300 mA/°C Temperature coefficient of Isc
beta -5.00 V/°C .. -0.001 V/°C Temperature coefficient of Uoc
gamma -5.00 %/°C .. 0.999 %/°C Temperature coefficient of Pmax
Rs 0.01 .. 9.99 Serial resistance of PV module
Selection of PV module type and parameters
Keys:
/ Selects appropriate option.
TEST Enters menu for changing type or parameters.
ESC, Function selector Exits back.
MEM Enters PV module type memory menu.
26


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3109 - Eurotest PV Lite at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3109 - Eurotest PV Lite in the language / languages: English as an attachment in your email.

The manual is 1.66 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3109 - Eurotest PV Lite

Metrel MI 3109 - Eurotest PV Lite Quick start guide - Dutch - 68 pages

Metrel MI 3109 - Eurotest PV Lite User Manual - Dutch - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info