The 4mm plug to BNC adapter accessory is for use on cable distribution systems and low voltage cables only. The operator must
verify that the circuit is safe for testing and take the appropriate precautions.
In the event of no reflections being visible, increase gain in setup mode until any reflection can be easily identified. (If no
reflections can be seen, try shorting or earthing the far end of the cable to ensure that you are “seeing” the whole length of the
cable).
The range may be changed by moving the four-way switch up and down and the cursor moved using the LEFT and RIGHT joy
switch. Move the cursor to the beginning of the reflection. The distance to the fault can then be directly read from the display.
The distance calculation is performed using the velocity factor set in the TDR. If this velocity factor is not correct for the cable
under test, the displayed distance will be incorrect.
Shown are two typical trace displays. The first shows an open circuit; the second a short circuit.
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Megger TDR1000-3 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Megger TDR1000-3 in the language / languages: English as an attachment in your email.

The manual is 2,49 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Megger TDR1000-3

Megger TDR1000-3 User Manual - German - 30 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info