Table I-1. Glucose TRUE Alert Performance using Calibration every 12 hours
mg/dL Insertion Site Glucose TRUE Alert Rate
Threshold Only Predictive Only Threshold & Predictive
30 min 15 min 30 min 15 min 30 min 15 min
50 Abdomen - - - - - -
Buttock 0.0% (0/1) 0.0% (0/1) 0.0% (0/1) 0.0% (0/1) 0.0% (0/2) 0.0% (0/2)
60 Abdomen - - 0.0% (0/1) 0.0% (0/1) 0.0% (0/1) 0.0% (0/1)
Buttock 100.0% (1/1) 100.0% (1/1) 25.0% (1/4) 25.0% (1/4) 40.0% (2/5) 40.0% (2/5)
70 Abdomen 0.0% (0/1) 0.0% (0/1) 0.0% (0/4) 0.0% (0/4) 0.0% (0/5) 0.0% (0/5)
Buttock 100.0% (2/2) 100.0% (2/2) 40.0% (2/5) 40.0% (2/5) 57.1% (4/7) 57.1% (4/7)
80 Abdomen 25.0% (1/4) 25.0% (1/4) 16.7% (1/6) 0.0% (0/6) 20.0% (2/10) 10.0% (1/10)
Buttock 40.0% (2/5) 40.0% (2/5) 33.3% (3/9) 22.2% (2/9) 35.7% (5/14) 28.6% (4/14)
90 Abdomen 50.0% (2/4) 50.0% (2/4) 50.0% (3/6) 33.3% (2/6) 50.0% (5/10) 40.0% (4/10)
Buttock 100.0% (8/8) 100.0% (8/8) 81.8% (9/11) 63.6% (7/11) 88.9% (16/18) 77.8% (14/18)
180 Abdomen 100.0% (7/7) 100.0% (7/7) 80.0% (8/10) 70.0% (7/10) 85.7% (12/14) 78.6% (11/14)
Buttock 89.7% (26/29) 89.7% (26/29) 84.8% (28/33) 81.8% (27/33) 85.7% (48/56) 83.9% (47/56)
220 Abdomen 100.0% (4/4) 100.0% (4/4) 50.0% (5/10) 50.0% (5/10) 61.5% (8/13) 61.5% (8/13)
Buttock 90.0% (18/20) 85.0% (17/20) 62.1% (18/29) 55.2% (16/29) 70.5% (31/44) 63.6% (28/44)
250 Abdomen 100.0% (5/5) 100.0% (5/5) 83.3% (5/6) 83.3% (5/6) 90.9% (10/11) 90.9% (10/11)
Buttock 80.0% (12/15) 73.3% (11/15) 65.2% (15/23) 56.5% (13/23) 70.3% (26/37) 62.2% (23/37)
300 Abdomen 60.0% (3/5) 60.0% (3/5) 40.0% (2/5) 40.0% (2/5) 50.0% (5/10) 50.0% (5/10)
Buttock 66.7% (4/6) 66.7% (4/6) 35.0% (7/20) 30.0% (6/20) 42.3% (11/26) 38.5% (10/26)
Note: Given the small sample size of data available in the low range, the alert performance in the 50 mg/dL and 60 mg/dL should be interpreted
with caution and may not reflect actual use performance.
Note: For the blank cells (-), there are no evaluable events in this reference range.
Glucose FALSE Alert Rate
The glucose false alert rate is the rate at which the blood glucose did not confirm
that the CGM alert was triggered correctly. For example:
False Threshold Hypoglycemic alert rate alerted when the CGM read that the
user was below the low threshold but the users blood glucose was actually above
that low threshold.
False Threshold Hyperglycemic alert rate alerted when the CGM read that the
user was above the high threshold but the users blood glucose was actually
below that high threshold.
False Predictive Hypoglycemic alert rate alerted when the CGM predicted that
the user would be below the low threshold but the user’s blood glucose was
actually above that low threshold within 15 or 30 minutes.
398 Chapter 17
416


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Medtronic Minimed 770G mgdl at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Medtronic Minimed 770G mgdl in the language / languages: English as an attachment in your email.

The manual is 32.39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Medtronic Minimed 770G mgdl

Medtronic Minimed 770G mgdl User Manual - German - 407 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info