Carb ratios
Maximum ratio settings Range
8
1 to 200 grams/unit
0.075 to 15 units/exch
Delivery accuracy
For a basal rate of 1.0 U/h, the delivery accuracy is ±5%.
For a basal rate of 0.025 U/h, the delivery accuracy is ±10%.
Delivery accuracy for bolus volumes < 0.1 unit is ±20% and delivery accuracy
for bolus volumes 0.1 unit is ±5%.
All Normal boluses are delivered within 16 minutes, 41 seconds ±3 seconds at
Standard rate (25 units, at 1.5 units per minute), and within 1 minute, 41
seconds ±3 seconds at Quick rate (25 units, at 15 units per minute).
The maximum infusion pressure generated and the occlusion threshold
pressure using a 3.0-mL reservoir is 90.67 kPa (13.15 psi). The average resulting
bolus volume generated upon clearing the occlusion is 0.0112 mL (equivalent
to 1.12 units of U-100 insulin).
The following is a representative delivery accuracy curve. The Trumpet Curve
represents the maximum percentage change from the expected insulin
dosage to the actual insulin delivered for a given time interval, known as the
observation window, during the infusion of insulin. The upper curve
corresponds to positive changes, and the lower curve corresponds to negative
changes.
256 Chapter 15
270


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Medtronic MiniMed 640G at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Medtronic MiniMed 640G in the language / languages: English as an attachment in your email.

The manual is 23.36 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Medtronic MiniMed 640G

Medtronic MiniMed 640G User Manual - German - 365 pages

Medtronic MiniMed 640G Additional guide - Dutch - 1 pages

Medtronic MiniMed 640G User Manual - Dutch - 331 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info