Electrostatic Discharge
Although your pump is designed to be unaected by typical levels of electrostatic
discharge (ESD), very high levels of ESD can result in a reset of the pump's
software with an associated pump error alarm. In most cases, exposure to high
levels of ESD will trigger the pump's A-13 alarm although, under certain
circumstances, high level ESD exposure can cause A-44, Bolus Stopped or Max
Delivery alarms. High levels of ESD are more likely in situations where the relative
humidity is very low, such as inside a heated building during the winter in areas
where it is cold outside.
If your pump experiences an A-13 or other error alarm, press the ESC and ACT
buttons to clear the alarm. If you are unable to clear the alarm by pressing ESC
and ACT, you may need to remove and replace the pump's battery to clear the
alarm. After clearing the alarm, you should always verify that your pump is set to
the correct date and time and that all other settings (basal rate, max basal and
bolus limits, etc.) are programmed to the desired values, since the software reset
could erase your previously programmed settings. Please see Alarms and alerts, on
page 199 for more details regarding what to do if your pump displays an error
alarm or other alert message.
Please contact our 24 Hour HelpLine at 1 800 646 4633 to report any error alarms or
other problems that occur with your pump.
WARNING: The Threshold Suspend feature will cause the pump to
temporarily suspend insulin delivery for two hours when the sensor
glucose reaches a set threshold. Under some conditions of use the
pump can suspend again resulting in very limited insulin delivery.
Prolonged suspension can increase the risk of serious hyperglycemia,
ketosis, and ketoacidosis. Before using the Threshold Suspend
feature, it is important to read the Threshold Suspend information in
the Getting Started Guide and the MiniMed 530G System User Guide
and discuss proper use of the Threshold Suspend feature with your
healthcare provider.
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Medtronic MiniMed 530G at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Medtronic MiniMed 530G in the language / languages: English as an attachment in your email.

The manual is 12.74 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info