5 Operating
5.1.3 Evaluating the control test
result
The permissible range for the control
solution reading is indicated on the
label of the test strip container. Your
test result must lie within the range
indicated. Make sure that you compare
the test result with the correct range.
If the control test result lies within the
range indicated on the test strip
container, then the monitor and the
test strips are working accurately.
If the control test result does not lie
within the range indicated on the test
strip container, then the following
options will be displayed to rectify the
problem:
Cause
Has the test strip been lying around
open for a long time?
Was the test strip container
properly closed?
Did the monitor work properly?
Is the control solution soiled or has it
exceeded the expiry date?
Were the test strips and control
solution stored in a cool, dry place?
Did you follow the steps of the test
procedure properly?
Remedy
Repeat the test with a test strip that has
been stored correctly.
The test strips are damp.
Replace the test strip.
Repeat the test as described in section
5.1.2. If problems persist, get in touch
with the service centre.
Use new control solution to check the
performance of the
monitor.
Repeat the control test using strips and
solution that have been stored
correctly.
Repeat the test as described in section
5.1.2. If problems persist, get in touch
with the service centre.
79026_MediTouch_GA/GB_09_01_12.qxd:Blutzucker-Messgerät 09.01.2012 13:46 Uhr Seite 17
21


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Medisana 79026 - MediTouch at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Medisana 79026 - MediTouch in the language / languages: English as an attachment in your email.

The manual is 2,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Medisana 79026 - MediTouch

Medisana 79026 - MediTouch User Manual - English, German, Dutch - 138 pages

Medisana 79026 - MediTouch User Manual - Dutch - 46 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info