EMC tables
Guidance and manufacturer's declaration-electromagnetic emission
The TS29 IR Thermometer is intended for use in the electromagnetic environment specied
below. The customer or the user of the TS29 IR Thermometer should assure that it is used in
such an environment.
Emissions test Compliance Electromagnetic environment guidance
RF emissions CISPR 11
Group 1 The TS29 IR Thermometer uses
RF energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The TS29 IR Thermometer is suitable
for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage uctuations/icker
emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer's declaration-electromagnetic immunity
The TS29 IR Thermometer is intended for use in the electromagnetic environment specied
below. The customer or the user of the TS29 IR Thermometer should assure that it is used in
such an environment.
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic environment-
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Power frequency
(50/60 Hz)
magnetic eld IEC
61000-4-8
3 A/m 3 A/m
Power frequency magnetic elds should
be at levels characteristic of a typical
location in a typical.
ENG
25
25


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Mediblink M300 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Mediblink M300 in the language / languages: English as an attachment in your email.

The manual is 0,39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info