531794
119
Zoom out
Zoom in
Previous page
1/124
Next page
Electromagnetic immunity
The electric breastpump Swing Maxi is intended for use in the electromagnetic environment
specied below. The customer or the user of the electric breastpump Swing Maxi should
assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2,5 GHz
10 Vrms
10 V/m
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the electric
breastpump Swing Maxi, including
cables, than the recommended
separation distance calculated from
the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance
d = 0,35P
d = 0,35P 80 MHz to 800 MHz
d = 0,7P 800 MHz to 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and d is the recommended sepa-
ration distance in metres (m).
Field strengths from xed RF trans-
mitters, as determined by an electro-
magnetic site survey,
a
should be less
than the compliance level in each fre-
quency range.
b
Interference may occur in the vicinity
of equipment marked with the follow-
ing symbol:
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 4)
119
119


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Medela Swing Maxi at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Medela Swing Maxi in the language / languages: English, Dutch, French as an attachment in your email.

The manual is 8,35 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info