NEBULIZER INSTRUCTIONS
Designed and created in accordance with the latest standards, is an Medical Device (Risk Class
IIa - MDD 93/42/EEC) highly efficient in medical therapies administered by means of aerosol
nebulization.
SAFETY WARNINGS
• A single-patient use is recommended
• The device should be used only with compressor nebulizers for aerosoltherapy that comply
with standards in force and whose use is indicated in the instruction manual
• Device not suitable for anaesthesia and lung ventilation
• The device should be used only with original accessories shown in this manual
Information concerning connection and use with devices for aerosoltherapy are shown in the
instruction manual of the devices. Always observe instructions and safety warnings as
indicated in the instruction manuals of the devices for aerosoltherapy
• Use the device and its accessories as per your doctor’s recommendations
• Use medications prescribed by your doctor according to his indications
• Do not use the device for other use than Nebulizer for Aerosoltherapy. Manufacturer will not
be held responsible for inappropriate use of the device
The device does not come in sterile pack; always carry out cleaning and sterilization operations
before and after each treatment. When using the device follow ordinary hygiene and cleaning
precautions, (wash your hands and carry out cleaning operations as described in the specific
section of this manual)
The device has small components which might be removed and easily swallowed. Use by
minors and disabled people require presence of an adult with his faculties. Do not leave the
device unattended in places easily accessible by minors and disabled people
• Always store the device in a dry and clean place far from light, heat and weather
• Dispose of device as provided by standards in force
In presence of pathologies with microbial contamination and infection hazard a singlepatient
use of accessories and nebulizer is recommended (always consult your doctor)
HOW TO PREPARE AND TO OPERATE THE DEVICE
The device is not sterile. Before use carry out cleaning and disinfection operations as described
in the specific section. To pour the medication rotate the nebulizer top counter clockwise and
remove it. Pour the quantity of medication prescribed by your doctor into the nebulizer tank.
IMPORTANT: In case you do so, empty the nebulizer, clean it thoroughly and proceed with new
filling. Once the medicinal product has been poured, screw again the top part on the bottom
one; then connect the air tube to the air outlet hub. Connect the other end of the tube to the
connection on the lower part of the nebuliser and then connect the accessories prescribed by
the doctor to the nebuliser itself.
Start the compressor for treatment to begin.
IMPORTANT: NEVER INHALE IN HORIZONTAL POSITION.
CLEANING INSTRUCTION
Before and after each treatment clean thoroughly each component of the nebulizer removing
medication residuals and possible impurities. Clean each component as described here below
except the air tube which -in case of treatment to a new patient or in case of impurities should
be replaced.
Clean all components with lukewarm water and/or neutral soap; rinse carefully removing excess
water and let them dry in the open air in a clean place.
To disinfect the accessories, proceed as follows:
1) Rotate the upper part of the nebuliser counter-clockwise.
2) Disconnect the pisper within the bottom of the nebuliser by pulling firmly with the fingers.
3) To disinfect the accessories, it is possible to use denatured spirit or a hypochlorite solution
that can be easily purchased at any chemist’s shop.
4) After use, disassemble the ampoule and clean all parts using lukewarm water; rinse carefully
and remove excess water with a soft cloth; then, let them dry in the open air in a clean place
Never boil the accessories.
USEFUL LIFE OF THE APPLIANCE
For each single patient, it is suggested to use the ampoule for 6 months or for 120 treatments
maximum. The nebuliser has to be replaced after a long standstill, if it shows deformations or
breaks, or in case the atomisation nozzle is clogged by dry medicinal product, dust, etc.
Use exclusively original accessories foreseen and indicated by the Manufacturer.
STORAGE AND TRANSPORTATION CONDITIONS
Storage and transportation temperature: +10°C (50°F) ÷ +40°C (104°F)
Storage and transportation humidity percentage: 30% ÷ 75% RH
OPERATING TECHNICAL DATA
Operating Minimum flow: > 5 l/min
Relative operating pressure: 72.4 kPa (0.724 bar) +/- 10%.
Min. capacity: 1 ml
Max. capacity: 8 ml
Read carefully instructions for use
15
Instructions and guarantee
GB
English
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Laica NE2007 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Laica NE2007 in the language / languages: English, German, French, Spanish as an attachment in your email.

The manual is 1,74 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info