38 39
COMPONENTS
1) LCD DISPLAY
2) “O/I” KEY
3) M KEY
4) ARMBAND
5) BATTERY COMPARTMENT
6) BATTERIES
7) CASE
TECHNICAL FEATURES
Pressure indication range: between 0 and 300 mmHg (arm cuff pressure)
between 30 and 180 pulsations/min. (pulse frequency)
Precision: ±3 mmHg (arm cuff pressure)
± 5% of the read value (pulse frequency)
Inflating and deflating: automatic
Power supply: 2 alkaline batteries type AAA (LR03)
Memories: 60 Memories per area (2 areas)
Working conditions: between +5°C and + 40°C; 85% relative humidity or lower
Storage conditions: between -10°C and + 50°C; 95% relative humidity or lower
Wrist circumference: between 135 and 215 mm approx.
Dimensions: 77x62x33 mm
Weight: about 120 g (without alkaline batteries)
Attention. Carefully check the instructions for use.
Type BF appliance
IP22: Degree of protection of electrical equipment covers, where the first figure indicates the
degree of protection against penetration by solid foreign bodies (from 0 to 6) and the
second figure the degree of protection against penetration by liquids (from 0 to 8).
ELECTROMAGNETIC COMPATIBILITY
This device complies with current regulations relating to electromagnetic compatibility (EMC)
Ensure that no liquids enter the appliance.
This appliance is not waterproof.
Place the device in the supplied case and, to store it, observe the environmental conditions
stated here below.
SPECIAL CASES
If the following circumstances occur:
1) After pressing the “O/I” key the reading does not start.
Check the batteries and replace them if needed.
Clean the contacts of the batteries with a dry cloth.
2) Nothing appears on the display when the “M” key is pressed.
The batteries are flat.
3) The batteries do not last long.
use alkaline batteries to increase working autonomy. If normal zinc-carbon batteries are
used they will need changing more often.
4) You cannot complete a reading.
check the batteries and change them if needed.
5) The blood pressure levels are different with each reading - the readings are very low or very
high.
the measuring position is not correct - Read the paragraph “Correct measuring position”
again.
the blood pressure readings vary constantly depending on the time of day and the state
of nerves of the person. Read the paragraphs “Warnings” and “Important” again.
6) The heart beat readings are too high or too low:
The person moved while the reading was being taken.
The reading was taken after physical exertion.
Read the paraghaphs “CORRECT MEASURING METHOD”, “WARNINGS” and
“IMPORTANT” again.
7) On the display, appears the symbol .
It has been detected the presence of an irregular pulse (arrhythmias). Repeat the
measurement. If the symbol appears again, refer to a physician to
establish the most suitable treatment to be followed.
Instructions and guarantee
EN
EnglishEnglish
EN
WRIST BLOOD PRESSURE MONITOR
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Laica BM1006 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Laica BM1006 in the language / languages: English, German, French, Italian, Portuguese, Spanish as an attachment in your email.

The manual is 1,62 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info