Front Panel Operation
2001-900-01 Rev. K/ August 2010 3-41
Trigger level
The frequency function has an adjustable trigger level. An
appropriate trigger level is needed for the frequency counter
to operate properly. The instrument only counts cycles with
peak amplitudes that reach the trigger level. For example, if
the trigger level is set for 5V, cycles with peak amplitudes
less than 5V are not counted.
If using AC+DC coupling, make sure the trigger level
accounts for the DC bias level. For example, if a 1V peak-to-
peak input signal is riding on a 5V DC bias level, a trigger
level of 5.5V is appropriate.
While the display is showing frequency measurements, the
RANGE and keys can be used to change the trigger
level of the measurement.
Each press of a RANGE or key adjusts the trigger level
by 0.5% of the presently selected maximum signal level to a
maximum of 60% of the range. The AUTO RANGE key
returns the trigger level to 0V or 0mA. After pressing one of
the range keys, the present trigger level is momentarily
displayed. The trigger level is also available as a multiple
display.
See Table 3-21 for the trigger level ranges and increments.
Frequency configuration
The following paragraphs detail how to change the Model
2001 from its bench reset conditions for frequency
measurements. The configuration menu is summarized in
Table 3-22. Note that a function does not have to be selected
in order to be configured. When the function is selected, it
will assume the programmed status.
NEXT
DISPLAY
PREV
POWER
DCV ACV DCI ACI Ω2 Ω4
FREQ TEMP
REL TRIG STORE RECALL
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
RANGE
AUTO
FILTER MATH
RANGE
2001 MULTIMETER
SENSE
Ω 4 WIRE
HI
INPUT
LO
INPUTS
CAL
500V
PEAK
F
R
FRONT/REAR
2A 250V
AMPS
350V
PEAK
1100V
PEAK
Model 2001
Caution : Maximum Input = 1100V peak, 2 x 10
7
V•Hz
AC Voltage
Source
A. AC Voltage Input
NEXT
DISPLAY
PREV
POWER
DCV ACV DCI ACI Ω2 Ω4
FREQ TEMP
REL TRIG STORE RECALL
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
RANGE
AUTO
FILTER MATH
RANGE
2001 MULTIMETER
SENSE
Ω 4 WIRE
HI
INPUT
LO
INPUTS
CAL
500V
PEAK
F
R
FRONT/REAR
2A 250V
AMPS
350V
PEAK
1100V
PEAK
Model 2001
Caution : Maximum Input = 1A peak
AC Current
Source
B. AC Current Input
+000.00 mHz
Terminals: VOLT Coupling: AC
+000.00 mHz
Terminals: CURR Coupling: AC
q~ÄäÉ=PJON
Trigger level range and increments
Maximum
signal level
Trigger level
Range Increment
1V
10V
100V
1000V
-0.600 to +0.600V
-6.00 to +6.00V
-60.0 to +60.0V
-600 to +600V
0.005V
0.05V
0.5V
5V
1mA
10mA
100mA
1A
-0.600mA to +0.600mA
-6.00mA to +6.00mA
-60.0mA to +60.0mA
-600mA to +600mA
0.005mA
0.05mA
0.5mA
5mA
Note: The trigger level for TTL triggering defaults to +0.80V.
73


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2001 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2001 in the language / languages: English as an attachment in your email.

The manual is 2,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info