Front Panel Operation
3-2 2001-900-01 Rev. K/ August 2010
WAR NI NG
The power cord supplied with the Model
2001 contains a separate ground wire for
use with grounded outlets. When proper
connections are made, instrument chassis
is connected to power line ground through
the ground wire in the power cord. Failure
to use a grounded outlet may result in per-
sonal injury or death due to electric shock.
3.2.2 Line fuse replacement
A rear panel fuse located below the AC receptacle protects
the power line input of the instrument. If the fuse needs to be
replaced, perform the following steps:
WAR NI NG
Make sure the instrument is disconnected
from the power line and other equipment
before replacing the line fuse.
1. With the power off, place the end of a flat-blade screw-
driver into the rear panel LINE FUSE holder. Push in
gently and rotate the fuse carrier one-quarter turn coun-
terclockwise. Release pressure on the holder and its in-
ternal spring will push the fuse carrier out of the holder.
2. Remove the fuse and replace it with the same type
(0.5A, 250V, slow blow, 5 20mm). The Keithley part
number is FU-71.
CAUTION
Do not use a fuse with a higher current
rating than specified, or instrument
damage may occur. If the instrument re-
peatedly blows fuses, locate and correct
the cause of the trouble before replacing
the fuse. See the optional Model 2001
Repair Manual for troubleshooting in-
formation.
3. Install the new fuse and fuse carrier into the holder by
reversing the above procedure.
3.2.3 Power-up sequence
To turn on the power, simply push in the front panel POWER
switch. Power is on when the switch is in the inner (1)
position. To turn off power, press POWER a second time to
release the switch.
On power-up, the Model 2001 performs self-tests on its
EPROM and RAM, and checksum tests on data stored in
non-volatile memory. (See Table 3-1.) If a failure is detected,
the instrument momentarily displays an error message and
the ERR annunciator turns on. (Messages are listed in Table
3-2.)
NOTE
If a problem develops while the
instrument is under warranty, return it to
Keithley Instruments, Inc. for repair (see
paragraph 1.3.).
If the instrument passes the self-tests, the firmware revision
levels, memory option (if installed), and presently selected
IEEE-488 primary address are displayed. An example of this
display is shown as follows:
Model 2001
Rev. A01 A01 MEM1 IEEE Addr=16
The firmware revision levels (left to right) are for the main
microcontroller and display microcontroller. The revision
level number may be different in your particular unit. If the
MEM1 or MEM2 memory options are not present, that
portion of the bottom line is left blank. The IEEE-488
address is its default value of 16.
Next, if the unit is configured to display the calibration due
date at power-up, the unit shows the following:
Model 2001
Calibration due: mmm/dd/yy
where “mmm” is the month abbreviation, “dd” is the day,
and “yy” is the year. If no calibration date is set, the display
shows that it is due now. (See the Model 2001 Calibration
Manual to set the calibration due date and paragraph 3.12.3
of this manual to set the display option.)
After the power-up sequence, the instrument begins its
normal display, for example:
000.0000 mVDC
Power-up error messages
Error messages that may be displayed during power-up are
summarized in Table 3-2. These are shown when one of the
checksum tests of Table 3-1 fails.
34


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2001 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2001 in the language / languages: English as an attachment in your email.

The manual is 2,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info