Table 5-2
CALCulate command summary
Command Description
Default
parameter
SCPI
:CALCulate[1]
:FORMat <name>
:FORMat?
:KMATh
:MMFactor <NRf>
:MMFactor?
:MBFactor <NRf>
:MBFactor?
:MUNits <name>
:MUNits?
:PERCent <NRf>
:ACQuire
:PERCent?
:STATe <b>
:STATe?
:DATA?
Subsystem to control CALC 1:
Select math format (NONE, MXB, PERCent).
Query math format.
Path to configure math calculations:
Set “m” factor for mx+b (-100e6 to 100e6).
Query “m” factor.
Set “b” factor for mx+b (-100e6 to 100e6).
Query “b” factor.
Specify units for mx+b reading (three characters ‘A’ through
‘Z’).
Query “mx+b” units.
Set target value for PERCent calculation (-100e6 to 100e6).
Use input signal as target value.
Query percent.
Enable or disable kmath calculation.
Query state of kmath function.
Read result of kmath calculation.
PERCent
1
0
MXB
1
(Note)
:CALCulate2
:FORMat <name>
:FORMat?
:STATe <b>
:STATe?
:IMMediate
:IMMediate?
:DATA?
Subsystem to control CALC 2:
Select math format: (MEAN, SDEViation, MAXimum,
MINimum, or NONE).
Query math format.
Enable or disable calculation.
Query state of math function.
Recalculate raw input data in buffer.
Perform calculation and read result.
Read math result of CALC 2.
NONE
(Note)
:CALCulate3
:LIMit[1]
:UPPer
[:DATA] <n>
[:DATA]?
:LOWer
[:DATA] <n>
[:DATA]?
:STATe <b>
:STATe?
:FAIL?
:CLEAR
[:IMMediate]
:AUTO <b>
:AUTO?
:IMMediate
Subsystem to control CALC 3 (limit test):
Path to control LIMIT 1 test:
Path to configure upper limit:
Set upper limit (-100e6 to 100e6).
Query upper limit.
Path to configure lower limit:
Set lower limit (-100e6 to 100e6).
Query lower limit.
Enable or disable limit test.
Query state of limit test.
Query test result (1 = pass, 0 = fail).
Path to clear failed test:
Clear failed test indication.
Enable or disable auto clear.
Query auto clear.
Re-perform limit tests.
1
-1
OFF
ON
*Note: ON is the *RST default parameter, and OFF is the :SYSTem:PRESet default.
2000-900 (J - Aug 2010) BOOK.fm Page 8 Wednesday, October 12, 2011 12:29 PM
139


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2000 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2000 in the language / languages: English as an attachment in your email.

The manual is 5,8 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Keithley 2000

Keithley 2000 User Manual - German - 122 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info