*STB? — Status Byte Query Read status byte register
Description
Use the *STB? query command to acquire the value (in decimal) of the Status Byte Reg-
ister. The Status Byte Register is shown in Figure 4-16. The binary equivalent of the decimal
value determines which bits in the register are set.
All bits, except Bit B6, in this register are set by other event registers and queues. Bit 6
sets when one or more enabled conditions occur.
The *STB? query command does not clear the status byte register. This register can only
be cleared by clearing the related registers and queues. See “status structure” for details.
For example, for an acquired decimal value of 48, the binary equivalent is 00110000. This
binary value indicates that bits 4 and 5 if the Status Byte Register are set.
The bits of the Status Byte Register are described as follows:
Bit 0, Measurement Status (MSB)-A set bit indicates that a measurement event has
occurred. The event can be identified by reading the Measurement Event Status
Register using the :STATus:MEASurement? command (see Section 5 for details).
Bit 1-Not used.
Bit 2, Error Available (EAV)-A set bit indicates that an error or status message is
present in the Error Queue. The message can be read using one of the following
SCPI commands:
:SYSTem:ERRor?
:STATus:QUEue?
See Section 5 for more information.
Bit 3, Questionable Summary Bit (QSB)-A set bit indicates that a calibration error has
occurred.
Bit 4, Message Available (MAV)-A set bit indicates that a message is present in the
Output Queue. The message is sent to the computer when the Model 2000 is ad-
dressed to talk.
Bit 5, Event Summary Bit (ESB)-A set bit indicates that an enabled standard event
has occurred. The event can be identified by reading the Standard Event Status Reg
-
ister using the *ESE? query command.
Bit 6, Master Summary Status (MSS)/Request Service (RQS)-A set bit indicates that
one or more enabled Status Byte conditions have occurred. Read the MSS bit by us
-
ing the STB? query command, or perform a serial poll to detect the occurrence of a
service request (RQS bit set).
Bit 7, Operation Summary (OSB)-A set bit indicates that an enabled operation event
has occurred. The event can be identified by reading the Operation Event Status
Register using the :STATus:OPERation? query command (see Section 5 for details).
2000-900 (J - Aug 2010) BOOK.fm Page 50 Wednesday, October 12, 2011 12:29 PM
129


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2000 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2000 in the language / languages: English as an attachment in your email.

The manual is 5,8 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Keithley 2000

Keithley 2000 User Manual - German - 122 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info