The following paragraphs discuss configuring the multimeter for making measurements.
See the end of Appendix A for information about optimizing readings for speed or accuracy.
The selected measurement range affects both the ultimate digits and accuracy of the
measurements as well as the maximum signal that can be measured. The range setting
(fixed or auto) for each measurement function is saved when changing functions.
The full scale readings for every range on each function are 20% overrange except for
the 1000VDC, 750VAC, 3ADC, 3AAC, and diode test ranges.
Input values more than the maximum readings cause the "OVERFLOW" messages to be
displayed.
To select a range, simply press the RANGE or key. The instrument changes one
range per keypress. The selected range is displayed for one second.
If the instrument displays the "OVERFLOW" message on a particular range, select a high-
er range until an on-range reading is displayed. Use the lowest range possible without caus-
ing an overflow to ensure best accuracy and resolution.
Note that the temperature and continuity functions have just one range.
To enable autoranging, press the AUTO key. The AUTO annunciator turns on when au-
toranging is selected. While autoranging is selected, the instrument automatically chooses
the best range to measure the applied signal. Autoranging should not be used when opti
-
mum speed is required.
Note that up-ranging occurs at 120% of range, while down-ranging occurs at 10% of nom-
inal range.
To cancel autoranging, press AUTO or the RANGE or key. Pressing AUTO to cancel
autoranging leaves the instrument on the present range.
The AUTO key has no effect on the temperature, continuity, and diode test functions.
Measurement Options 3-3
2000-900 (J - Aug 2010) BOOK.fm Page 3 Wednesday, October 12, 2011 12:29 PM
50


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2000 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2000 in the language / languages: English as an attachment in your email.

The manual is 5,8 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Keithley 2000

Keithley 2000 User Manual - German - 122 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info