Measuring frequency and period
The Model 2000 can make frequency measurements from 3Hz to 500kHz on voltage
ranges of 100mV, 1V, 10V, 100V, and 750V. Period measurements can be taken from 2µs
to 333ms on the same voltage ranges as the frequency.
The instrument uses the volts input terminals to measure frequency. The AC voltage
range can be changed with the RANGE
and keys. The signal voltage must be greater
than 10% of the full-scale range.
CAUTION The voltage limit is subject to the 8 × 10
7
V•Hz product.
Trigger level
Frequency and Period use a zero-crossing trigger, meaning that a count is taken when
the frequency crosses the zero level. The Model 2000 uses a reciprocal counting technique
to measure frequency and period. This method generates constant measurement resolution
for any input frequency. The multimeter’s AC voltage measurement section performs input
signal conditioning.
2000-900 (J - Aug 2010) BOOK.fm Page 26 Wednesday, October 12, 2011 12:29 PM
37


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2000 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2000 in the language / languages: English as an attachment in your email.

The manual is 5,8 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Keithley 2000

Keithley 2000 User Manual - German - 122 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info