Measuring voltage
The Model 2000 can make DCV measurements from 0.1µV to 1000V and ACV measure-
ments from 0.1µV to 750V RMS, 1000V peak.
Connections
Assuming factory default conditions, the basic procedure is as follows:
1. Connect test leads to the INPUT HI and LO terminals. Either the front or rear inputs
can be used; place the INPUTS button in the appropriate position.
2. Select the measurement function by pressing DCV or ACV.
Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE and keys to select
a measurement range consistent with the expected voltage.
4. Connect test leads to the source as shown in Figure 2-4.
CAUTION Do not apply more than 1000V peak to the input or instrument damage may
occur. The voltage limit is subject to the 8 × 10
7
V•Hz product.
5. Observe the display. If the “OVERFLOW” message is displayed, select a higher
range until an o normal reading is displayed (or press AUTO for autoranging). Use
the lowest possible range for the best resolution.
6. Take readings from the display.
Crest factor
AC voltage and current accuracies are affected by the crest factor of the waveform, the
ratio of the peak value to the RMS value. Table 2-3 lists the fundamental frequencies at
which the corresponding crest factor must be taken into account for accuracy calculations.
Table 2-3
Crest factor limitations
Crest factor Fundamental frequency
2
3
4-5
50kHz
3kHz
1kHz
2000-900 (J - Aug 2010) BOOK.fm Page 18 Wednesday, October 12, 2011 12:29 PM
29


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2000 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2000 in the language / languages: English as an attachment in your email.

The manual is 5,8 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Keithley 2000

Keithley 2000 User Manual - German - 122 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info