33
appear, the measurment begins. During ination
the device already determines measured values
used to estimate the required ination pressure. If
this pressure is insufcient, the device automatically
inates by another 40 mmHg (Real Fuzzy Logic).
• Then the pressure in the cuff is slowly released and
the pulse is measured.
• When the measurement has been completed, the
remaining air pressure is released very quickly. The
pulse, the systolic and the diastolic blood pressure
are displayed.
• The device requires approx. 5 seconds to store the
determined measured value.
• To switch off and release the pressure, press the
Power button again. If you forget to switch off the
device, it switches off automatically after approx. 1
minute.
Wait at least 5 minutes before taking another
measurement!
The computer automatically saves the blood pressure
readings of the last 50 measurements. When the 50
memory locations are exceeded, the current oldest
value is deleted.
• Press the user-switching key to select a
memory zone and press the M-key to switch from
8 | DISPLAY OF THE SAVED READINGS
memory to memory with sequence numbers.
• The saved measured values are continuously stored.
• To switch off, press the -key again.
• If you forget to switch off the computer, it will
switch off automatically after 1 minute.
Error messages can occur when:
• the blood pressure value is unusually high or low (EE
appears in the display);
• you move or talk during the measurement (EE
appears in the display);
• the cuff tube is not properly inserted (E1 appears in
the display);
• ination takes longer than 20 seconds (E1 appears in
the display);
• the ination pressure is higher than 300 mmHg (E2
appears in the display);
• an error occurs when storing the measured values
(E3 appears in the display).
In the about cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and
that you do not move or talk. Put the batteries back in
if necessary or else replace them.
9 | TROUBLESHOOTING
33


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Inventum BC26 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Inventum BC26 in the language / languages: English, German, Dutch, French as an attachment in your email.

The manual is 0,85 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info