783862
9
Zoom out
Zoom in
Previous page
1/16
Next page
14 15
Control Solution
Control solution contains a known amount of glucose that
reacts with test strips and is used to ensure that your meter
and test strips are working together properly.
Material needed:
• iHealth Wireless Smart Gluco-Monitoring System (BG5)
• iHealth AGS-1000I Blood Glucose Test Strip
• Control Solution
Perform Control Solution test when:
• First receiving the meter
• Checking the meter and test strips at least once a week
• Using a new vial of test strips
You suspect the meter or test strips are not working properly
Your blood glucose test results are not consistent with how
you feel, or if you think the results are not accurate
• Practicing the testing process
• The meter has been dropped or damaged
Step 1: Insert the test strip to turn on the meter
Insert the test strip in the meter’s Strip Port, and wait for the
Blood Sample Symbol to appear in the meter display.
Step 2: Press the Memory Button to
turn on the control (“CTL”) mode.
When the “CTL” symbol appears, the
meter is in control ode and will not save
this test results in memory.
Performing an Offline Control Solution Test with Meter
Alone (not using iOS device)
When taking a Control Solution Test with meter alone, it is
important to set the test mode to “CTL,” so that the result
will not be mixed up with your regular test results. Control
Solution Test results will not be saved in the meter’s memory.
Please follow the instructions below.
Step 3: Press the “Memory Button” again to turn off the
“CTL” mode and switch back to the regular testing
mode.
When the “CTL” symbol disappears from the meter display,
the meter is in regular testing mode and subsequent tests
will not be treated as a control solution test.
Step 4: Apply Control Solution
• Shake the Control Solution vial before each use.
• Squeeze a drop of Control Solution and place it in the vial
cap. For better results, it is recommended that you use the
second drop of the Control Solution (discard the first drop).
• Hold the meter and move the absorbent hole of the test
strip to catch the drop. Once the confirmation window fills
completely, the meter will start counting down. Please
leave the test strip blood samples when the countdown
begins.
Note: To avoid contaminating the entire vial of Control
Solution, do not directly apply Control Solution onto a strip.
Step 2: Tap the “CTL” button to switch to Control (“CTL”)
mode
Step 1: Launch the App
I Performing an Online Control Solution Test (with iOS
device)
Step 5: Read and Compare the Results
After counting down to “1” the Control
Solution test result will appear on the
meter display.
The result of the Control Solution test
should be within the range printed on the
test strip vial label. If the test result falls
outside the specified range, repeat the test, carefully
following the steps above.
9


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for iHealth BG5 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of iHealth BG5 in the language / languages: English as an attachment in your email.

The manual is 0.89 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of iHealth BG5

iHealth BG5 User Manual - German - 35 pages

iHealth BG5 User Manual - French - 17 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info