506529
15
Zoom out
Zoom in
Previous page
1/30
Next page
27 28
Power Source:
Four 1.5V DC (AA) batteries
Measurement
Method:
Oscillometric
Measurement Range:
Pressure: 40~280 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute
Accuracy:
Pressure: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Manual Inflation
Deflation:
Automatic Pressure release valve
Memory Capacity:
60 memories
Auto-shut-off:
1 minute after last button operation
Operation
Environment:
Temperature: 50°F~104°F (10°C ~40°C)
Humidity: 15% – 90% R.H.
Storage Environment:
Temperature: -4°F~158°F (-20°C~70°C)
Humidity: Less than 90% R.H.
Weight:
.43 lbs/193 grams (without batteries)
Arm Circumference:
Standard size adult arm cuff:
9”–13” (23 – 33 cm)
Dimensions:
5”(L) x 4.1”(W) x 2.7”(H)
126mm(L) x 105mm(W) x 68mm(H)
Internally Powered equipment
Type BF Equipment
IPx0-No special protection against the
ingress of water
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or
nitrous oxide.
Continuous operation with short-time loading.
SPECIFICATIONS
Note: These specifications are subject to change without notice.
TROUBLESHOOTING
If any abnormality arises during use, please check the following points
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn
on when the START/
STOP
button is pushed.
Batteries have
run down.
Replace them with
two new AA alkaline
batteries.
Battery polarities
have been positioned
incorrectly.
Re-insert the batteries
in the correct
positions.
EE measurement
error symbol shown
on display or the
blood pressure
value is displayed
excessively low
(or high).
Cuff has been placed
incorrectly.
Wrap the cuff properly
so that it is positioned
correctly.
Did you talk or move
during measurement?
Measure again.
Keep arm steady
during measurement.
Shaking of the arm
with the cuff on.
E1 error symbol
shown on display
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Check cuff connection.
Measure again.
E2 error symbol
shown on display
Inflation pressure
exceeding 300 mmHg.
Turn the unit off, then
measure again.
E3 error symbol
shown on display
Error determining
measurement data.
Measure again.
BPS-060_B.indd 29-30BPS-060_B.indd 29-30 5/01/10 1:27 PM5/01/10 1:27 PM
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Homedics BPS-060 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Homedics BPS-060 in the language / languages: English, Spanish as an attachment in your email.

The manual is 2,49 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info