Inserting and changing batteries
1. Push out the tray at the bottom of the blood pressure monitor so that the battery
compartment door is exposed.
2. Press slightly on the handle of door while you lift and remove it. Insert or change the batteries
in the battery compartment. Make sure that the batteries are placed correctly. The door is put
back in its place.
3. The batteries must be changed when the battery symbol on the display shows low battery.
The battery symbol will disappear once the device is switched on and the battery is charged.
As the supplied batteries are for test only, they may be discharged earlier than batteries you buy
in stores. Replace the batteries in pairs. Remove batteries when unit is not in use for extended
periods of time.
Error search
If irregularities occur during use, please check the following items before you contact your vendor:
If the device still does not function, please contact the store where the device was purchased.
Do not attempt to open the device or adjust any of the inner parts yourself.
Establishing arrhythmia (heart rhythm disturbances)
An unstable heart rhythm can lead to or indicate cardiovascular diseases. By means of
advanced technology, this blood pressure device can warn the user that he or she has an
irregular heart rhythm which in turn contributes to better health monitoring.
If the symbol
IHB
appears during blood pressure measurement, it means that an irregular heart
rhythm has been established. This should only be regarded as a warning. Note that it is
important that the user is relaxed, sits still, and does not speak during measurement. It is
strongly recommended that you contact your doctor if the irregular heart rhythm symbol appears
several times in succession (5 times or more).
Memory
This blood pressure device can be used by two users. Each user can store up to 60 measurements.
1. Press .
2. Choose user 1 or 2 by pressing .
3. Press to enter memory.
4. The blood pressure device will first show the average of the three most recent measurements
stored in the memory.
5. Every time is pressed, previously stored measurements are shown. The latest
measurement is shown first.
6. To clear memory press to choose user 1 or 2.
7. Press and hold the button for approximately 5 seconds and all data for the chosen user
are erased automatically.
Recalibration service
All blood pressure devices must be recalibrated to ensure continual measurement accuracy.
The blood pressure device should be recalibrated after 2 years of normal use (approx. 3
measurements a day). Subsequently, the device will show CA on the display. The device should
also be recalibrated if e.g. it has been dropped on the floor or in fluids, if fluid has been spilled over
or on the device, or if it has been exposed to extremely high or low temperatures.
This monitor must only be calibrated by authorised personel and should therefore be sent to
HoMedics Group Ltd for re-calibration. Do not carry out any repairs yourself. Please contact the
shop where you bought the product for further details on the procedure for submitting the monitor or
contact HoMedics customer service on +44 (0) 1732 378557 or email: cservice@homedics.co.uk.
10
GB
GB
Symptom Possible reasons Correction
The device does not switch
on when the stop/start button
is activated.
A change of batteries is required. Change to four new batteries.
EE measurement error
symbol in the display, or
the blood pressure
measurement is excessively
low (or high).
The cuff is fitted incorrectly.
Fit the cuff again so that it is
positioned correctly.
Did you talk or move during
measurement?
Restart measurement and keep
the arm steady during the entire
measurement.
Shaking of the arm while wearing
the cuff.
E1 error symbol on the
display.
Airflow is irregular. The tube is not
properly applied to the device.
Check the cuff’s connection. Restart
measurement.
E2 error symbol on the
display.
The inflation pressure exceeds 300mmHg. Switch off the device and restart
measurement.
E3 error symbol on the
display.
Data error. Remove and reinsert batteries.
CA on the display. Calibration warning. See page 48 for more information.
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Homedics BPA-200 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Homedics BPA-200 in the language / languages: English as an attachment in your email.

The manual is 3,43 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info