64
• Close the bottle tightly after use.
• Left over control solution should not be added back into the
control bottle.
• Store the bottle of smartLAB
®
nG control solution at room
temperature, below 30°C (86°F). Do not freeze nor refrigerate.
• If you would like to purchase smartLAB
®
nG control solutions,
please contact your local dealer.
Running a smartLAB
®
control solution test
Please make sure you have all necessary items at hand when run-
ning a control solution test. You need the smartLAB
®
global W nG
meter, a smartLAB
®
nG test strip, and smartLAB
®
nG control solu-
tion.
1. Switch on the meter
Take a strip out of the vial and immediately close it. Insert the strip
into the meter in the direction of the arrow on the strip. The meter
turns on automatically and the symbol for applying blood.
2. Running the control solution test
When the symbol for applying blood appears, shortly shake the
bottle with the smartLAB
®
nG
control solution. Open the bottle and
wipe the tip of the bottle with a tissue. Squeeze a little drop of con-
trol solution on a clean, not adsorbing surface (plastic, glass, etc.)
andclose the bottle immediately. Hold the application/reaction
zone of the test strip vertically onto the drop of control solution.
The control solution is automatically sucked in the reaction zone of
the test strip. The signal tone indicates that enough blood entered
the reaction zone. The meter will now start with the measurement