154 GMC-I Messtechnik GmbH
Maintenance – Calibration
11.6 Displays – Error Messages
Displays and Error Messages Meaning / Cause Remedy
LCD does not light up
Mains power cable not plugged in Plug in the power cable (see page 16).
Batteries are excessively discharged
Leave test system connected to the mains with the
power cable for approx. 30 min.
ONSTART was not pressed
Press and hold the ONSTART key for at least
3 seconds (see page 19).
LCD no longer reacts when touch screen is activated Press the reset key (see page 14).
Probe test The probe must be connected in order to contact
the housing which is connected to the protective
conductor.
Connect probe 1/2. Connect the probe to sockets 1 and 2 (see page 14).
Switch the device under test on.
The DUT connected to the test socket must be
switched on for insulation testing.
Switch the device under test on.
Caution: Execute all safety measurements before
starting the function test.
Safety precaution
Do not start the function test until after the protective
conductor resistance measurement for safety class 1,
the insulation resistance measurement and the leak-
age current measurements have been executed and
passed by the device under test.
Error (PE potential test)
Mains connection error: voltage at the protective
conductor (PE) when the contact surface is touched
with the finger.
Error (PE connection test)
The test system’s mains power cable is defective. Inspect the mains power cable.
Mains connection error: no PE contact. Check PE contact at mains power supply.
Error (probe test) The probe is not connected. Connect the probe to sockets 1 and 2 (see page 14).
Error (short-circuit test) Short-circuit at the test socket
There is a short-circuit in the device under test or in
the DUT connector cable.
OFF (on test) The device under test is not switched on. Switch the device under test on.
154


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUSTAR FM at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUSTAR FM in the language / languages: English as an attachment in your email.

The manual is 1,75 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info