22 GMC-I Messtechnik GmbH
Initial Start-Up – Setup
4.7 General Information Regarding Data Backups
There are a total of three ways to back up data:
The backup function (see section 4.11.3)
Data export to XML (see section 7.7.6)
Data export to the L&H format (see section 7.7.6)
The backup function creates a stored image of all database
entries and measurement results. This function has the
disadvantage of a version-dependent backup, i.e. the data can
only be restored to the same version of the test system from which
they were backed up. Backups which have been created before
software updates are thus no longer usable..
All data are saved when export to an XML file is executed. The
XML file can also be read after updating the test system, and can
be imported in a trouble-free fashion. For this reason, you should
always export your data to an XML file before installing an update.
In the event of data export to the L&H format, the data bases in
PS3 and in the test system are reconciled. In this case, only data
(catalog / data) which are important for the PS3 database are
exported. This type of data export cannot be used as a substitute
for a backup, or for data export to an XML file.
Note
“Data export to XML” is preferable to the “backup function”
for assuring secure, long-term data archiving.
Attention!
!
If the backup is interrupted due to a reset or by switching
the instrument off, the database may be destroyed.
CF Card (data safety)
To ensure data safety, the CF card must always remain in the in-
strument, and the auto-backup function must be activated. This is
the only way of saving backup copies of the batabase during the
power-down process.
4.8 Icons in the Main Menu Footers
Icon Designation Meaning
Description
Setup Menu setup, system settings
See section
4.9
Accessories Accessory programs, e.g. pocket calculator
See section
4.11.11
Man Test Menu for individual tests and manual test See section 5
Auto-Test Menu for automated test sequences See section 6
Admini-
stration
Menu for administration of DUTs,
customers etc.
See section 7
Back Button Go back one menu level
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUSTAR FM at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUSTAR FM in the language / languages: English as an attachment in your email.

The manual is 1,75 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info