68 GMC-I Messtechnik GmbH
22 Maintenance - Recalibration
22.1 Housing Maintenance
No special maintenance is required for the housing. Keep outside
surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the
use of cleansers, abrasives or solvents.
22.2 Recalibration
The respective measuring task and the stress to which your measuring in-
strument is subjected affect the ageing of the components and may result
in deviations from the guaranteed accuracy.
If high measuring accuracy is required and the instrument is frequently
used in field applications, combined with transport stress and great tem-
perature fluctuations, we recommend a relatively short calibration interval
of 1 year. If your measuring instrument is mainly used in the laboratory and
indoors without being exposed to any major climatic or mechanical stress,
a calibration interval of 2-3 years is usually sufficient.
During recalibration* in an accredited calibration laboratory
(DIN EN ISO/IEC 17025) the deviations of your instrument in relation to
traceable standards are measured and documented. The deviations de-
termined in the process are used for correction of the readings during
subsequent application.
We are pleased to perform DAkkS or factory calibrations for you in our cal-
ibration laboratory. Please visit our website at www.gossenmetrawatt.com
( Company DAkkS Calibration Center or FAQs Calibration
questions and answers).
By having your measuring instrument calibrated regularly, you fulfill the re-
quirements of a quality management system per DIN EN ISO 9001.
Standards DIN VDE 0701-0702 and IEC 63353 (VDE 0751) stipulate that
only measuring instruments which are regularly tested and calibrated may
be used for testing.
* Verification of specifications or adjustment services are not part of the calibration.
For products from our factory, however, any necessary adjustment is frequently
performed and the observance of the relevant specification is confirmed.
22.3 Safety Checks
Check the safety of your test instrument on a regular basis. We recom-
mend the same test intervals as for recalibration.
The SECUTEST... is designed as a totally insulated instrument in accor-
dance with standards IEC 61010 and VDE 0404. The protective conduc-
tor is only used for measurement purposes and is therefore not accessible
in idle state. A protective conductor test at the test socket can be per-
formed as follows:
Ð Connect the SECUTEST... to a multiple distribution box.
Ð Conduct a contact current measurement for permanently installed
DUTs (the test socket may not be connected to any load).
Ð Measure the protective conductor resistance between the neighbour-
ing socket of the multiple distribution box and the test socket.
Ð The measured value may not exceed 0.3 Ω.
Insulation resistance between LN and PE in the SECUTEST... equals ap-
prox. 150 kΩ for metrological reasons.
This must be taken into account when performing the safety checks, i.e.
the protective conductor current measurement must produce a value
below 3.5 mA, instead of insulation resistance measurement (if the equiv-
alent leakage current measurement method has been chosen, the value
must remain below 7 mA).
Apart from this, the SECUTEST... features 3 exposed conductive parts for
which contact current measurement must produce a value below 0.5 mA:
RS232 interface
metallized start key
protective conductor clip at the test socket.
68


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info