Measured value P and calculated value S are compared, and the smaller value is displayed.
Reference Ranges
Line Voltage 115 / 230 V ±0.2%
Line Frequency 50/60 Hz ±0.1%
Waveshape sine (deviation between effective and
rectified value < 0.5%)
Ambient Temperature +23 °C ±2K
Atmospheric Humidity 50% relative ±5%
Load Resistors linear
Nominal Ranges of Use
Line Voltage 103.5 V 126.5 V or 207 V 253 V
Line Frequency 50 Hz or 60 Hz
Line Voltage Waveshape sine
Temperature 0 °C + 50 °C
Func-
tion
Measured Quantity Measuring Range /
Nominal Range of
Use
Resolu-
tion
Open-
Circuit
Voltage
U
0
Short-
Circuit
Current
I
K
Internal
Resis-
tance
R
I
Measuring
Uncertainty
8
Intrinsic
Uncertainty
8
Overload Capacity
Value Duration
Functions Test
Nominal Voltage U
LN
103,5 V ... 126,5 V
207.0 ... 253.0 V
0.1 V ±(2.5%rdg.+5 digits) 253 V cont.
Load Current I
a
0 ... 16.00 A
RMS
10 mA ±(2.5%rdg.+5 digits) 20 A 10 min
Active Power P 0 ... 3700 W
10
1W
±(5% rdg.+10 digits)
>20digits
253 V cont.
20 A 10 min
Apparent Power AP 0 ... 4000 VA 1 VA Calculated Value U
LN
• I
V
±(5% rdg.+10 digits)
>20digits
Power Factor PF,
sinusoidal: cos
ϕ
0.00 ... 1.00 0.01 Calculated Value P / AP, Display > 10 W ±(10% rdg.+5 digits)
Residual Current
ΔI
between L and N
0.00 ... 31.00 mA 10 μA—
±(10% rdg.+10 d)
> 10 digits
±(5% rdg.+5 digits)
11
U
AC/DC
Voltage
0 ... 253.0 V
, and
0.1 V ±(5% rdg.+10 d)
±(2.5%rdg.+5 digits)
>10digits
253 V cont.
U
Probe
Probe Voltage
0 ... 253.0 V
, and
0.1 V
±(2.5%rdg.+5 digits)
>10digits
253 V cont.
R Resistance 0 ... 150.0 kΩ 100 Ω <20V– 1.1mA
±(1% rdg.+3 digits) 253 V cont.
I
CLIP
Current via
Clip-On Current-
Voltage Converter
WZ12C
0.000 ... 10.00 A
1mA 1.5MΩ
±(3% rdg.+10 digits)
>10digits
without clip
253 V cont.
0 ... 100 A
1A 1.5MΩ 253 V cont.
Temp
Temperatur e
with Pt100 / Pt1000
Sensor
– 200 ... – 50
°C1°C
<20V– 1.1mA
±(2% rdg.+1 °C) 10 V cont.
– 50.1 ... + 300.0
°C0.1°C ±(1% rdg.+1 °C) 10 V cont.
+300 ... +850
°C1°C ±(2% rdg.+1 °C) 10 V cont.
61


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info