590774
57
Zoom out
Zoom in
Previous page
1/72
Next page
GMC-I Messtechnik GmbH 57
Visual inspection
OK?
SC I: RPE
OK?
Short-circuit
at DUT?
Test
OK?
Yes/No
No
Display results, save / print report
Yes
No
Yes
No
No
Framed with dashed line:
The test is only run if it has been activat-
ed in the initial program window, or in
the Setup menu under Sequence...
.
Switch line voltage to test socket, start function test
Start
function test
Run all
tests?
Contact all exposed, conductive parts with the probe.
The test can be repeated as often as desired for various protective
conductor parts with the manual sequence
1)
1)
If it is not clear whether or not all
exposed, conductive parts are
connected to one another or to the
protective conductor, testing can be
performed in the manual mode.
No
Set test conditions:
1. Potential conductor interrupted
2. Operational earth interrupted
3. Connect housing to protective conductor
4. Connect application parts to PE
Earth leakage current, contact current
2)
, patient leakage current
3)
, patient auxiliary current
4)
Mains poles
reversed?
No
Reverse L and N
Set single-fault conditions:
1. N interrupted
5)
2. Normal condition
3. Protective conductor interrupted
4. Mains to application part
5)
Run
all SFC?
Yes
Yes
No
2)
no N interrupted
3)
Test is only
performed if at least
one type BF or CF
application part is
connected.
4)
Test is only
performed if at least
two type BF or CF
application parts are
connected
5)
Test method: equiv-
alent leakage current
R-INS
OK?
Yes
Yes/No
Switch DUT on
Activate high-voltage
Yes
HV OK?
No
Test HV?
Additional insulation resistance measurement between application parts an PE
Select connection, select test regulation EN 60601 classify DUT (SC I, II or III), test conditions (amongst others R
INS
measurement), application part? (type B/BF/CF)
Prüfung
O.K.?
57


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email