15.7 Testing Extension Cables for VDE 0701-0702 (VDE 0701 Part 1)
(optional EL1 adapter)
Extension Cables up to 5 m Long
Protective conductor resistance between the earthing contact at the
mains plug and all exposed metal parts may not exceed 0.3 Ω for safety
class I devices. This value may not exceed 1 Ω for permanently con-
nected data processing systems (DIN VDE 0701, part 240).
Extension and Connector Cables Longer than 5 m
An additional cable resistance of 0.1 Ω, however, not more than 1 Ω, is al-
lowable as of a length of 5 m for each additional 7.5 m.
Resistance testing for cables longer than 5 m is thus advisable (see also
limit values on page 20.
Note
+
The EL1 accessory adapter is required for testing for short-
circuiting and interruption of single-phase extension cables.
Connecting the Extension Cable or the Multiple Outlet
Note
+
The keys in the adapter’s handle have no function.
Performing the Test
õ Connect the extension cable to the EL1 adapter as shown in the figure
above.
õ Select the “
Ext. Cable
WITH EL1
” test from the initial menu with the cursor
and acknowledge with the key:
x
.
õ Select “Start Testing” with the key.
õ Start the test sequence with the key.
õ
First perform and acknowledge visual inspection of the extension cable.
õ Enter the length of the cable with the
and keys. Acknowledge with
the key.
For conductors with a rated current of
> 16 A, the cross section (Diameter)
has to be taken into account during
measurement.
The cross section is only queried if a 3-
phase current adapter has been connec-
ted and the connection parameter has
been set to VL2E or AT3-IIIE. Upon ente-
ring the cross section, the limit value for
the protective conductor resistance is
recalculated.
PROBE 5 4
(21)
EL1
SECUTEST
Device Under Test =
Extension Cable
Test Socket
46


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info