590775
7
Zoom out
Zoom in
Previous page
1/72
Next page
GMC-I Messtechnik GmbH 7
1.2 Table: Individual Measurements and Regulations
Key
Standards printed in grey half-tone will be superseded by the new
DIN VDE 0701-0702:2008 standard.
Required test
1.3 Table: Leakage Current Types
Key
NC = normal condition
PAP = patient application part
PE = Potential earthing
, system protective conductor
PC = Protective conductor of the DUT
Individual Measurements
per Regulation
Test Current [A]
DIN VDE 0701-0702
DIN VDE 0701
DIN VDE 0701
DIN VDE 0701
DINEN60950
DINEN61010
DINEN60335
IEC 62353
IEC 601/EN 60 601 2nd
IEC 601/EN 60 601 3rd
Protective Conductor
Resistance
0.2
••••
10
25
••••••
Insulation Resistance
••
Equivalent Leakage Current
••••
High-Voltage Test
••• AC AC
Equivalent (Device) Leakage
Current
••
Equivalent Patient Leakage Current
Residual Current
••
Contact Current
•• ••• ••
Earth Leakage Current
••
Patient Leakage Current
•••
Total Patient Leakage Current
Patient Auxiliary Current
••
Device Leakage Current
Single Fault Conditions N
PE
Mains at Application Part
••
DIN VDE
0701-0702
IEC 62353
(VDE 0751-1)
DIN EN 60601-1 The following is
measured:
Equivalent
Leakage Current
PROBE
(connected to
protective
conductor)
to L & N
Equivalent
Device Leakage
Current
I
EL
interrupted
from N
PROBE
(connected to
protective
conductor)
to L & N
Equivalent
Patient Leakage
Current
L & N & PE to
Patient Jacks
Contact Current
Contact Current
NC
Probe to PE
Leakage Current
from applied part
Patient Leakage
Current
NC
Patient Jack to PE
Patient Auxiliary
Current
NC
Patient Jack to
Patient Jack
Earth Leakage
Current
NC
Protective
Conductor to PE
Device Leakage
Current during
Operation, Direct
Measurement
Protective
Conductor
Interrupted, Probe
+ PAP to PE
Protective
Conductor
Current with
Residual
Current
Measurement
Device Leakage
Current during
Operation,
with residual
Current
Measurement
See chapter 11.5
7


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info