GMC-I Messtechnik GmbH 7
Safety Warnings
2 Safety Features and Precautions
This instrument fulfills the requirements of applicable European
and national EC directives. This is confirmed by means of the CE
mark. A corresponding declaration of conformity can be
requested from GMC-I Messtechnik GmbH.
The SECULIFESR measuring instrument has been manufactured
and tested in accordance with the following safety regulations:
IEC 61010-1 / DIN EN 61010-1 / VDE 0411-1, DIN VDE 0404
IEC 61577 / EN 61577 / VDE 0413 part 1, 2 and 3
When used for its intended purpose, the safety of the user, the
measuring instrument and the device under test (electrical
equipment or electrical medical device) is assured.
Read the operating instructions carefully and completely before placing
your measuring instrument into service. Follow all instructions contained
therein. Make sure that the operating instructions are available to all
users of the instrument.
Tests may only be performed by a qualified electrician, or under
the supervision and direction of a qualified electrician. The user
must be instructed by a qualified electrician in the execution and
evaluation of tests.
Note
Manufacturers and importers of electrical medical devices
must provide documentation for the performance of
maintenance by trained personnel.
Observe the following safety precautions:
The instrument may only be connected to electrical supply
systems with 230 V/240 V which conform to the valid safety
regulations (e.g. IEC 60364, VDE 0100) and are protected with
a fuse or circuit breaker with a maximum rating of 16 A.
Measurements within electrical systems are prohibited.
Be prepared for the occurrence of unexpected voltages at
devices under test (for example, capacitors may be
dangerously charged).
Make certain that the measurement cables are in flawless
condition, e.g. no damage to insulation, no cracks in cables or
plugs etc.
Insulation Resistance Measurement (alternative leakage current):
Testing is conducted with up to 500 V. Current limiting is
utilized (I < 10 mA), but if the terminals (L and N) are touched,
electrical shock may occur which could result in consequential
accidents.
Leakage Current Measurement
It is absolutely essential to assure that the device under test is
operated with line voltage during performance of leakage
current measurements. Exposed conductive parts may
conduct dangerous contact voltage during testing, and may
not under any circumstances be touched (mains power is
disconnected if leakage current exceeds approx. 10 mA).
Function Test
Attention!
!
The function test may only be performed after the DUT has
successfully passed the safety test!
7


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE SR at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE SR in the language / languages: English as an attachment in your email.

The manual is 0,42 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info