54 GMC-I Messtechnik GmbH
Technical Data
8 Technical Data
1)
Values of less than 200 digits are suppressed in the mV range.
2)
At 0 ° ... + 40 °C
3)
Display up to max. 5.1 V, “OL” in excess of 5.1 V.
4)
Residual value deviates within 1 ... 30 d from the zero point due to TRMS converter
when probe tips are short-circuited. See frequency influence on page 56.
5)
Switch-off time/Cool-down time > 30 min and T
A
40 °C
Meas.
Function
Measuring Range
Resolution at Upper
Range Limit
Input Impedance
Intrinsic Errorunder Reference Conditions
Overload Capacity
2)
±(... % rdg. + ... d) ±(... % rdg. + ... d) ±(... % rdg. + ... d)
11,999 1199
~ /
~
4)
4)
Value Time
V
100 mV 10 μV 9MΩ≥ 9MΩ // < 50 pF 0.09 + 5 with ZERO 1 + 30 (> 300 d)
1)
1 + 30 (> 300 d)
1)
1000 V
DC
AC
RMS
sine
continuous
1 V 100 μV 9MΩ≥ 9MΩ // < 50 pF 0.05 + 3 0.5 + 9 (> 200 d) 1 + 30 (> 300 d)
10 V 1 mV 9MΩ≥ 9MΩ // < 50 pF 0.05 + 3 0.5 + 9 (> 200 d) 1 + 30 (> 300 d)
100 V 10 mV 9MΩ≥ 9MΩ // < 50 pF 0.05 + 3 0.5 + 9 (> 200 d) 1 + 30 (> 300 d)
1000 V 100 mV 9MΩ≥ 9MΩ // < 50 pF 0.09 + 3 0.5 + 9 (> 200 d) 1 + 30 (> 300 d)
Voltage drop, approx. at upper range limit
~
4)
4)
A
X-
TRA
OUTDOOR
PRO
100 μA 10 nA 12 mV 12 mV 0.5 + 5 1.5 + 10 (> 200 d) 1.5 + 30 (> 200 d)
0.2 A
continuous
1 mA 100 nA 120 mV 120 mV 0.5 + 3 1.5 + 10 (> 200 d) 1.5 + 30 (> 200 d)
10 mA 1 μA 16 mV 16 mV 0.5 + 3 1.5 + 10 (> 200 d) 1.5 + 30 (> 200 d)
100 mA 10 μA 160 mV 160 mV 0.5 + 3 1.5 + 10 (> 200 d) 1.5 + 30 (> 200 d)
1 A 100 μA 40 mV 40 mV 0.9 + 10 1.5 + 10 (> 200 d) 1.5 + 30 (> 200 d)
10 A: 5 min
5)
16 A: 30 s
5)
10 A 1 mA 600 mV 600 mV 0.9 + 10 1.5 + 10 (> 200 d) 1.5 + 30 (> 200 d)
A
T
ECH
10 mA 1 μA 16 mV 16 mV 0.1 + 5 1 + 10 (> 200 d) 1.5 + 30 (> 200 d)
0.2 A
continuous
100 mA 10 μA 160 mV 160 mV 0.1 + 5 1 + 10 (> 200 d) 1.5 + 30 (> 200 d)
1 A 100 μA 40 mV 40 mV 0.9 + 10 1 + 10 (> 200 d) 1.5 + 30 (> 200 d)
10 A: 5 min
5)
16 A: 30 s
5)
10 A 1 mA 600 mV 600 mV 0.9 + 10 1 + 10 (> 200 d) 1.5 + 30 (> 200 d)
Factor: 1:1/10/100/1000
Input Input impedance
A
T
ECH
0.1/1/10/100 A 100 mA
Current measuring input
(A socket)
Specification see current ranges A (T
ECH)
Measuring input
1/10/100/1000 A 1 A 0.2 A continuous
10 A: 5 min
10/100 / 1 00 0 / 10000 A
10 A plus clip-on current sensor error
A
T
ECH
BASE
0.1/1/10/100 A 100 mV
Voltage measurement input
T
ECH: (V socket) Ri =1 MΩ/9 MΩ
B
ASE: ( V socket) Ri ~1 MΩ
±(0.5% rdg. + 10 d)
±(1% rdg. + 30 d)
> 300 d
±(1% rdg. + 30 d)
> 300 d
Measurement input
1/10/100/1000 A 1 V
1000 V RMS Max. 10 s
10/100 / 1 00 0 / 10000 A
10 V Plus clip-on current sensor error
Open-circuit voltage
Meas. curr. @ range limit
±(... % rdg. + ... d)
Ω
100 Ω 10 mΩ < 1.4 V Approx. 300 μA 0.2 + 5
with active ZERO function
1000 V
DC
AC
RMS
sine
Max. 10 s
1kΩ 100 mΩ < 1.4 V Approx. 250 μA 0.2 + 5
10 kΩ 1 Ω < 1.4 V Approx. 100 μA 0.2 + 5
100 kΩ 10 Ω < 1.4 V Approx. 12 μA 0.2 + 5
1MΩ 100 Ω < 1.4 V Approx. 1.2 μA 0.2 + 5
10 MΩ 1kΩ < 1.4 V Approx. 125 nA 0.5 + 10
40 MΩ 10 kΩ < 1.4 V Approx. 20 nA 2.0 + 10
100 Ω —0.1Ω Approx. 8 V Approx. 1 mA const. 3 + 5
5.1 V
3)
1 mV Approx. 8 V Approx. 1 mA const. 0.5 + 3
X
-
TRA
/O
UTDOOR
PRO
TECH
54


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE HIT at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE HIT in the language / languages: English as an attachment in your email.

The manual is 0,93 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt SECULIFE HIT

Gossen Metrawatt SECULIFE HIT User Manual - Dutch - 70 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info